P6139: Difference between revisions

Jump to navigation Jump to search
m (adding "tds3052")
(adding accessory set 020-1956-01)
Line 17: Line 17:
Unless they've had their cables replaced, they had a whisker- or dendrite-growth problem in the cable that will short out the signal and drive you nuts.  It can be cleared by running some current through the BNC connector but only temporarily.  The only fix is to order a new cable from Tektronix, which I'm not sure you can still do.
Unless they've had their cables replaced, they had a whisker- or dendrite-growth problem in the cable that will short out the signal and drive you nuts.  It can be cleared by running some current through the BNC connector but only temporarily.  The only fix is to order a new cable from Tektronix, which I'm not sure you can still do.
</blockquote>
</blockquote>
==Accessories==
* [https://w140.com/tekwiki/wiki/020-1956-01 020-1956-01 Compact Probe Accessory Set]


==Manuals==
==Manuals==