The '''Tektronix 067-0587-02''' is a signal standardizer (calibration) plug-in for [[7000-series scopes]].
The '''Tektronix 067-0587-02''' is a signal standardizer (calibration) plug-in for [[7000-series scopes]].
It is a complete redesign of the [[067-0587-01]] standardizer.
It is a complete redesign of the [[067-0587-01]] standardizer. It provides a calibrated staircase at 1 line/Div in Gain mode, a fast-rising pulse in Step Resp mode,


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Risetime | 150 ps}}
{{Spec | |      — ''GAIN mode'' — }}
{{Spec | Bandwidth | 1 GHz}}
{{Spec | Staircase accuracy | ± 0.3% in ± 3 Div, ± 0.5% in ± 4 Div, Linearity ± 0.2% of 10 Div }}
{{Spec |  |      — ''STEP RESP mode'' — }}
{{Spec | Amplitude | 2 to 10 Div }}
{{Spec | Position  | Top at least ±5 Div from center of screen }}
{{Spec | Rise time | < 150 ps}}
{{Spec | Repetition rate | 10 Hz to 1 MHz in decade steps, 0.1 %}}
{{Spec | Pretrig Out | 0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable) }}
{{Spec |  | &nbsp;&nbsp;&nbsp;&nbsp;&nbsp;— ''FREQ RESP mode'' — }}
{{Spec | Input impedance | 50 Ω }}
{{Spec | Input range    | 0.5 to 1 V<sub>p-p</sub> }}
{{Spec | Flatness | ± 3% from 3 MHz to 1 GHz }}
{{EndSpecs}}
{{EndSpecs}}
==Internals==
The 067-0587-02 is based on a number of Tektronix custom ICs from the 1 GHz [[7104]] scope generation, such as the [[155-0160-00|155-0160-00 trigger amplifier + source select chip]] and the [[155-0175-00|155-0175-00 trigger amplifier chip]].
The mode selector switch is a special elastomer-based cam switch.  The manual warns that while it can be cleaned if proper procedure is followed (note some screws are labelled "do not remove" on the cover!), only at the factory can (could ...) it be repaired/re-assembled completely.


==Pictures==
==Pictures==
<gallery>
<gallery>
File:067-0587-00.jpg|067-0587-00
File:Tek-067-0587-02 front.jpg  | 067-0587-02 front view
File:Tek-067-0587-02 left.jpg  | 067-0587-02 left side
File:Tek-067-0587-02 right.jpg | 067-0587-02 right side
</gallery>
</gallery>


Line 26: Line 44:
[[Category:Specifications needed]]
[[Category:Specifications needed]]
[[Category:Calibration fixtures]]
[[Category:Calibration fixtures]]
[[Category:Introduced in 1979]]