10A1: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 11: | Line 11: | ||
}} | }} | ||
The '''Tektronix 10A1''' is a differential amplifier plug-in for [[647]] scopes. The input amplifier uses [[8056]] nuvistor | The '''Tektronix 10A1''' is a differential amplifier plug-in for [[647]] scopes. | ||
triodes for the input cathode followers. | The input amplifier uses [[8056]] nuvistor triodes for the input cathode followers. | ||
The 10A1 is a comparator, meaning that it has a built-in precision variable DC voltage source | |||
that can be added or subtracted from the signal voltage. | |||
==Specifications== | ==Specifications== | ||
Line 21: | Line 23: | ||
* CMRR below 100 kHz is 20,000:1 | * CMRR below 100 kHz is 20,000:1 | ||
* Maximum input current: less than 1 nA | * Maximum input current: less than 1 nA | ||
* Comparison voltage range: -6 V to +6 V | |||
* Comparison voltage accuracy (0C to 40C): +/- (0.1% + 5 mV) | |||
Revision as of 18:31, 30 April 2016
The Tektronix 10A1 is a differential amplifier plug-in for 647 scopes. The input amplifier uses 8056 nuvistor triodes for the input cathode followers. The 10A1 is a comparator, meaning that it has a built-in precision variable DC voltage source that can be added or subtracted from the signal voltage.
Specifications
- Maximum calibrated sensitivity: 1 mV/div
- Minimum calibrated sensitivity: 20 V/div
- Input impedance: 1 megohm in parallel with 20 pF
- Frequency response is DC to 45 MHz, except in maximum sensitivity setting, where it is 35 MHz
- CMRR below 100 kHz is 20,000:1
- Maximum input current: less than 1 nA
- Comparison voltage range: -6 V to +6 V
- Comparison voltage accuracy (0C to 40C): +/- (0.1% + 5 mV)
Pictures
-
-
-
-
-
-
-
Input Amplifier Schematic