Line 8: Line 8:
series=[[500-series scopes]]|
series=[[500-series scopes]]|
manuals=
manuals=
* [http://w140.com/kurt/tek_1s2_sn1990-up.pdf 1S2 manual (PDF, sn 1990 and up)]
* [[Media:070-0889-00.pdf|1S2 manual (PDF, sn 1990 and up, OCR)]]
* [http://www.pa4tim.nl/wp-content/uploads/2012/01/1S2calibrationPA4TIM.pdf 1S2 Calibration Guide by Fred Schneider, PA4TIM]
* [http://www.pa4tim.nl/wp-content/uploads/2012/01/1S2calibrationPA4TIM.pdf 1S2 Calibration Guide by Fred Schneider, PA4TIM]
* [http://www.pa4tim.nl/?p=3016 1S2 Page on the Website of Fred Schneider, PA4TIM]
* [http://www.pa4tim.nl/?p=3016 1S2 Page on the Website of Fred Schneider, PA4TIM]
* [[Media:Tek 1s2 fcp june 1966.pdf‎|Tektronix 1S2 Factory Calibration Procedure, June 1966]]
* [[Media:Tek 1s2 fcp aug 1968.pdf‎|Tektronix 1S2 Factory Calibration Procedure, August 1968]]
<!-- * [[Media:1S2_calibration_guide.pdf|1S2 Calibration Guide by Fred Schneider, PA4TIM (PDF, OCR)]] -->
}}
}}
The '''Tektronix 1S2''' is a [[sampling oscilloscope|sampling]]/TDR plug-in for [[500-series scopes]].  
The '''Tektronix 1S2''' is a [[sampling oscilloscope|sampling]]/TDR plug-in for [[500-series scopes]], [[introduced in 1967]] and sold through 1972.
 
Like the [[1S1]], it is a complete sampling subsystem and uses the mainframe merely as a power supply and low-speed X-Y display.
Like the [[1S1]], it is a complete sampling subsystem and uses the mainframe merely as a power supply and low-speed X-Y display.
The 1S2 outputs the Y-axis signal on plug-in interface connector pins 1 and 3 like other 500-series plug-ins.
The 1S2 outputs the Y-axis signal on plug-in interface connector pins 1 and 3 like other 500-series plug-ins, and the X-axis signal on the front panel, via a banana connector. The X-axis sensitivity of the oscilloscope mainframe is expected to be 1 V/div.
The 1S2 outputs the X-axis signal on the front panel, via a banana connector.
The X-axis sensitivity of the oscilloscope mainframe is expected to be 1 V/div.


Unlike the 1S1, it contains pulsers that allows it to perform time-domain reflectometry (TDR) measurements.   
Unlike the 1S1, the 1S2 contains pulsers that allows it to perform time-domain reflectometry (TDR) measurements.   
The 1S2 has two 50 Ω pulsers, one of which is active at a time.  One puts out a 250 mV pulse with 50 ps rise time.   
The 1S2 has two 50 Ω pulsers, one of which is active at a time.  One puts out a 250 mV pulse with 50 ps rise time.   
The other puts out a slower 1 V pulse.  The sampling bridge is a feed-through type.   
The other puts out a slower 1 V pulse.  The sampling bridge is a feed-through type.   


For TDR measurements, the signal from one of the internal pulsers is fed through the sampler, and out to the the transmission line being tested.   
For TDR measurements, the signal from one of the internal pulsers is fed through the sampler, and out to the transmission line being tested.   
Returning pulses are detected as they pass back through the sampler, and they are reabsorbed by the 50 Ω impedance of the pulser.   
Returning pulses are detected as they pass back through the sampler, and they are reabsorbed by the 50 Ω impedance of the pulser.   


Waveform output can be to the CRT of the mainframe and/or to a plotter or low-speed data acquisition device via banana connectors on the front of the 1S2.   
Waveform output can be to the CRT of the mainframe and/or to a plotter or low-speed data acquisition device via banana connectors on the front of the 1S2.   


The 1S2 was [[introduced in 1967]] and was sold through 1972.
The interface between the GR-874 connector and the printed circuit board is described in US Patent 3,426,311.
 
{{MissingSpecs}}


The interface between the GR-874 connector and the printed circuit board is described in US Patent 3,426,311.
==Internals==
The 1S2 contains an internal power supply producing regulated +19 V, -19 V, and -136 V outputs,
derived from a power transformer whose primary voltage is the 6.3 V<sub>AC</sub>
provided by the 500-series mainframe on pins 13 and 14 of the plug-in interface connector.
The regulators use bipolar transistors for the feedback amplifiers and the output emitter-followers, and zener diodes as references.


The 1S2 contains an internal power supply producing regulated +19 V, -19 V, and -136 V outputs.
As part of this internal power supply, the 1S2 has a power transformer whose primary voltage is the 6.3 V AC
provided my the 500-series mainframe on pins 13 and 14 of the plug-in interface connector.
The regulators use bipolar transistors for the feedback amplifiers and the output emitter-followers.
The references for the 1S2's internal power supplies are zener diodes in the 1S2.
==Links==
==Links==


Line 59: Line 64:
Tek 1s2 e 7.jpg
Tek 1s2 e 7.jpg
Tek 1s2 regulated ps.png|Internal Regulated Power Supply
Tek 1s2 regulated ps.png|Internal Regulated Power Supply
Tek 1s2 rear.jpg|rear
</gallery>
</gallery>


Line 66: Line 72:
[[Category:Time-domain reflectometers]]
[[Category:Time-domain reflectometers]]
[[Category:500 series plugins]]
[[Category:500 series plugins]]
[[Category:GR874]]

Revision as of 14:32, 8 June 2019

Template:Plugin Sidebar 2 The Tektronix 1S2 is a sampling/TDR plug-in for 500-series scopes, introduced in 1967 and sold through 1972.

Like the 1S1, it is a complete sampling subsystem and uses the mainframe merely as a power supply and low-speed X-Y display. The 1S2 outputs the Y-axis signal on plug-in interface connector pins 1 and 3 like other 500-series plug-ins, and the X-axis signal on the front panel, via a banana connector. The X-axis sensitivity of the oscilloscope mainframe is expected to be 1 V/div.

Unlike the 1S1, the 1S2 contains pulsers that allows it to perform time-domain reflectometry (TDR) measurements. The 1S2 has two 50 Ω pulsers, one of which is active at a time. One puts out a 250 mV pulse with 50 ps rise time. The other puts out a slower 1 V pulse. The sampling bridge is a feed-through type.

For TDR measurements, the signal from one of the internal pulsers is fed through the sampler, and out to the transmission line being tested. Returning pulses are detected as they pass back through the sampler, and they are reabsorbed by the 50 Ω impedance of the pulser.

Waveform output can be to the CRT of the mainframe and/or to a plotter or low-speed data acquisition device via banana connectors on the front of the 1S2.

The interface between the GR-874 connector and the printed circuit board is described in US Patent 3,426,311.

Key Specifications

  • please add

Internals

The 1S2 contains an internal power supply producing regulated +19 V, -19 V, and -136 V outputs, derived from a power transformer whose primary voltage is the 6.3 VAC provided by the 500-series mainframe on pins 13 and 14 of the plug-in interface connector. The regulators use bipolar transistors for the feedback amplifiers and the output emitter-followers, and zener diodes as references.

Links

Pictures