==Specifications==
==Specifications==
===Digital Multimeter===
{{BeginSpecs}}
{{Spec | DC Volts | 0.1 V to 1000 V,  Input Resistance: 10MΩ <br />Accuracy,  0.1v & 1v:  ±0.1%, ±3 counts; 10v & 100v: ±0.15%, ±1 count; 1000v: ±0.2%, ±1 count}}
{{Spec | True RMS Volts |
:  0.1v: ±2.5% (DC); ±1.5% (40 Hz – 4KHz); ±3.5% (4KHz – 40KHz)
:  1, 10, 100v:  ±2% (DC); ±1% (40Hz – 40KHz)
: 1000v:  ±2% (DC); ±1% (40 Hz – 4KHz); ±2% (4KHz – 40KHz)
}}
{{Spec | Current range input resistance | 0.1mA: 1000Ω; 1mA: 100Ω; 10mA: 10.2Ω; 100mA: 1.2Ω; 1000mA: 0.3Ω }}
{{Spec | DC Current |
:  0.1mA:  ±0.5%, ±3 counts; 1mA – 1000mA: ±0.25%, ±3 counts
}}
{{Spec | True RMS Current |
:  0.1mA:  ±2.5%, ±5 counts (DC); ±1.5%, ±5 counts (40Hz – 4KHz); ±4.5%,±5 counts (4KHz - 40KHz)
:1mA – 1000mA: ±2.5%, ±5 counts (DC); ±1.5%, ±5 counts (40Hz – 4KHz); ±3.5%,±5 counts (4KHz - 40KHz)
}}
{{Spec | Resistance |
:    1KΩ: ±0.5%, ±3 counts; 10KΩ – 1MΩ:  ±0.5%, ±1 count; 10MΩ: ±1%, ±1 count
}}
{{EndSpecs}}


DIGITAL MULTIMETER
===Oscilloscope===


DIRECT CURRENT VOLTS
{{BeginSpecs}}
 
{{Spec | Voltage ranges | 5 mV/Div to  – 100 V/Div, ±3% }}
    Range:  0.1v to 1000v
{{Spec | Bandwidth (Voltage) | DC – 1 MHz (at 10 mV/Div and below, 400 kHz) }}
 
{{Spec | Current ranges | 5 µA/Div to 100 mA/Div, ±3% }}
    Input Resistance: 10MΩ
{{Spec | Bandwidth (Current) | DC – 200 kHz (5 µA/Div 10 µA/Div); DC – 400 kHz (20 µA/Div 200 mA/Div) }}
 
{{Spec | Input resistance | 10 MΩ // 150 pF (5 mV/Div 1 V/Div), 100 pF (2 V/Div to 100 V/DIv) }}
    Accuracy, 0.1v & 1v:  ±0.1%, ±3 counts; 10v & 100v: ±0.15%, ±1 count; 1000v: ±0.2%, ±1 count
{{Spec | Rise time |  875 ns (5mV – 10mV/DIV); 350 ns (20mV – 100v/DIV) }}
 
{{Spec | Sweep |  500 ms/Div 2 µs/Div, ±5% (magnified or unmagnified) }}
TRUE RMS VOLTS
{{Spec | Batteries |  2 &times; “D” NiCd cells }}
 
{{Spec | Operating time | 3.5 hours typical at maximum intensity after full charge cycle}}
    0.1v: ±2.5% (DC); ±1.5% (40 Hz 4KHz); ±3.5% (4KHz – 40KHz)
{{Spec | Charge time | 16 h}}
 
{{Spec | Mains power | 90 – 136 V<sub>AC</sub>. 48 – 62 Hz (Option 1: 180 – 250 V<sub>DC</sub>. or 180 – 250 V<sub>AC</sub>, 48 – 62 Hz) }}
    1, 10, 100v:  ±2% (DC); ±1% (40Hz – 40KHz)
{{Spec | Power consumption | Less than 8 watts }}
 
{{EndSpecs}}
    1000v:  ±2% (DC); ±1% (40 Hz – 4KHz); ±2% (4KHz – 40KHz)
 
 
DC & RMS CURRENT
 
0.1mA = 1000Ωs; 1mA = 100Ω; 10mA = 10.2Ω; 100mA = 1.2Ω; 1000mA = 0.3Ω
 
DC CURRENT
 
    0.1mA:  ±0.5%, ±3 counts; 1mA – 1000mA: ±0.25%, ±3 counts
 
TRUE RMS CURRENT
 
    0.1mA:  ±2.5%, ±5 counts (DC); ±1.5%, ±5 counts (40Hz – 4KHz); ±4.5%,±5 counts (4KHz - 40KHz)  
 
    1mA – 1000mA: ±2.5%, ±5 counts (DC); ±1.5%, ±5 counts (40Hz 4KHz); ±3.5%,±5 counts (4KHz - 40KHz)
 
RESISTANCE
 
    1KΩ: ±0.5%, ±3 counts; 10KΩ – 1MΩ:  ±0.5%, ±1 count; 10MΩ: ±1%, ±1 count
 
 
 
OSCILLOSCOPE
 
VERTICAL:
 
    5mv – 100V/DIV, ±3%, DC 400KHz (5mV – 10mV/DIV); DC – 1MHz (20mV 100v/DIV
 
    5µA – 100mA/DIV, ±3%, DC – 200KHz (5µA – 10µA/DIV); DC 400KHz (20µA – 200mA/DIV)
 
Input resistance: 
 
    10MΩ, 150pF (5mV – 1V/DIV); 100pf (2V 100V/DIV)
 
Risetime:
 
    875ns (5mV – 10mV/DIV); 350ns (20mV – 100v/DIV)
 
HORIZONTAL:
 
SWEEP:
    500ms 2µs/DIV, ±5% (magnified or unmagnified)
 
BATTERIES:
 
    2 X “D” NiCd cells
 
    Typical operating time as maximum intensity after full charge cycle:  3.5 hours
 
    Charge time16 hours
 
MAINS OPERATION:
 
    90 – 136vac, 48 – 62 Hz (Option 1: 180 – 250 vdc or 180 – 250 vac, 48 – 62 Hz)
 
    Power consumption: Less than 8 watts


==Links==
==Links==

Revision as of 14:06, 16 September 2014

{{{manufacturer}}} 
Miniature portable 1 MHz oscilloscope and TRMS digital multimeter
Tektronix 213, front view

Produced from 1975 to 1988

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 213 is a miniature portable oscilloscope introduced in 1975. It combines a 3½-digit DMM and a 1 MHz single channel scope in one unit. The maximum sweep rate is 0.4 μs/div with ×10 sweep magnifier.

The 213 provides true RMS voltage and current measurements. It is battery or AC line powered.

Specifications

Digital Multimeter

Key Specifications

DC Volts 0.1 V to 1000 V, Input Resistance: 10MΩ
Accuracy, 0.1v & 1v: ±0.1%, ±3 counts; 10v & 100v: ±0.15%, ±1 count; 1000v: ±0.2%, ±1 count
True RMS Volts
0.1v: ±2.5% (DC); ±1.5% (40 Hz – 4KHz); ±3.5% (4KHz – 40KHz)
1, 10, 100v: ±2% (DC); ±1% (40Hz – 40KHz)
1000v: ±2% (DC); ±1% (40 Hz – 4KHz); ±2% (4KHz – 40KHz)
Current range input resistance 0.1mA: 1000Ω; 1mA: 100Ω; 10mA: 10.2Ω; 100mA: 1.2Ω; 1000mA: 0.3Ω
DC Current
0.1mA: ±0.5%, ±3 counts; 1mA – 1000mA: ±0.25%, ±3 counts
True RMS Current
0.1mA: ±2.5%, ±5 counts (DC); ±1.5%, ±5 counts (40Hz – 4KHz); ±4.5%,±5 counts (4KHz - 40KHz)
1mA – 1000mA: ±2.5%, ±5 counts (DC); ±1.5%, ±5 counts (40Hz – 4KHz); ±3.5%,±5 counts (4KHz - 40KHz)
Resistance
1KΩ: ±0.5%, ±3 counts; 10KΩ – 1MΩ: ±0.5%, ±1 count; 10MΩ: ±1%, ±1 count

Oscilloscope

Key Specifications

Voltage ranges 5 mV/Div to – 100 V/Div, ±3%
Bandwidth (Voltage) DC – 1 MHz (at 10 mV/Div and below, 400 kHz)
Current ranges 5 µA/Div to 100 mA/Div, ±3%
Bandwidth (Current) DC – 200 kHz (5 µA/Div – 10 µA/Div); DC – 400 kHz (20 µA/Div – 200 mA/Div)
Input resistance 10 MΩ // 150 pF (5 mV/Div – 1 V/Div), 100 pF (2 V/Div to – 100 V/DIv)
Rise time 875 ns (5mV – 10mV/DIV); 350 ns (20mV – 100v/DIV)
Sweep 500 ms/Div – 2 µs/Div, ±5% (magnified or unmagnified)
Batteries 2 × “D” NiCd cells
Operating time 3.5 hours typical at maximum intensity after full charge cycle
Charge time 16 h
Mains power 90 – 136 VAC. 48 – 62 Hz (Option 1: 180 – 250 VDC. or 180 – 250 VAC, 48 – 62 Hz)
Power consumption Less than 8 watts

Links

Pictures