2440

From TekWiki
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Manuals – Specifications – Links – Pictures

The Tektronix 2440 is a two-channel 300 MHz, 500 MS/s digital sampling oscilloscope from the late 1980s.

It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CCD output signals at 4 MS/s.

WARNING:
Operation of the 2440, 2432, 2432A, 2430, and 2430A: A/D-converter and CCD hybrids may suffer irreparable damage within a few minutes if operated out of the metal enclosure without extra cooling.

Key Specifications

Bandwidth 300 MHz (Limit: 20, 100, Full)
Rise time 1.2 ns (Repeat on, 16×averaging)
Sampling rate 500 MSa/s each channel with simultaneous acquisition
Sampling type Real time: 100 ns/Div and slower, Equivalent time 50 ns/Div and faster
Single event bandwidth 200 MHz
Deflection 2 mV/div to 5 V/div in 1-2-5 sequence
Resolution 8 bit
Record length 1024 samples
Auto scale factor On-Screen Readout indicates probe tip deflection factors for ×1, ×10, ×100 and ×1000 probes
Input impedance 1 MΩ // 15 pF or 50Ω
Sweep speeds 2 ns/div to 5 s/div in 1-2-5 sequence (Modes A, A Intensified and B)
Auto measure Simultaneously measures up to four parameters on one or more displayed waveforms, updated at 3 Hz intervals or display a snapshot of 20 parameters of waveform target
Acquisition modes Normal, Envelope (glitch capture), Average
Display modes CH1, CH2, invert, add, multiply, X-Y
Cursors Volts, Time, Volts At Time, 1/Time, Slope
Functions A Mode - Auto Level, Auto/Roll, Normal and Single Seq; B Mode - Trigger After Delay, Runs After Delay, 1K Record Length and more
Interface GPIB
Probes Two passive P6137 400 MHz ×10 voltage probes, 1.3 m
Power 90 to 132 VAC or 80 to 250 VAC, 48 to 440 Hz; 160 W
Features
  • Autoprobe Function With P6137 Probe
  • 300 MHz X-Y bandwidth at probe tip
  • Automatic measurements
  • extensive triggering capabilities
  • Printer/Plotter Output
  • On-screen HELP text describing every front-panel control
Options

Links

Internals

The 2440 employs several processors − an MC68B09 main CPU controls the instrument, a custom 16-bit waveform processor performs signal computations, and an MC6805 microcontroller handles the front panel interface. The GPIB Interface is implemented with a TMS9914 controller.

The power supply consists of a buck-type pre-regulator followed by a push-pull inverter and linear post-regulators (except for the +5 V digital supply rail).

CRT is a 154-0850-01.

Pictures

Components

Some Parts Used in the 2440

Part Part Number(s) Class Description Used in
148-0140-00 148-0140-00 Discrete component miniature bi-stable relay actuator 2424L 2430 2431L 2432 2439 2440 119-1445-02
154-0850-01 154-0850-00 154-0850-01 CRT CRT 2424L 2430 2430A 2431L 2432 2439 2440 2445 2465 2465A 2465B
155-0242-00 155-0242-00 155-0242-01 Hybrid integrated circuit Z-axis amplifier and autofocus 2215 2252 2440 2445 2445B 2465 2465B 2467B
155-0294-00 155-0294-00 155-0294-01 Monolithic integrated circuit Z-axis amplifier 2424L 2440
165-2273-00 165-2273-00 Hybrid integrated circuit peak detector 2432 2432A 2440
Motorola 6809 156-1494-00 156-1494-01 Monolithic integrated circuit 8-bit microprocessor 2424L 2440 7D20 DM5120 DM5520
TMS9914 156-1444-00 156-1444-01 Monolithic integrated circuit GPIB controller 2221 2230 2424L 2440 4052A 4054A 021-0374-00 067-1137-99 DC5009 DC5010 DM5010 DM5120 DM5520 MI5010 SG5010 SG5030 SI5010 PS5004 PS5010