2710

From TekWiki
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Manuals – Specifications – Links – Pictures

The Tektronix 2710 is a spectrum analyzer with a frequency range of 10 kHz to 1.8 GHz.

Key Specifications

Frequency 10 kHz to 1.8 GHz
Frequency Span 10 kHz/div to 100 MHz/div in 1–2–5 sequence, plus 0 Hz and MAX: 180 MHz/div
Resolution Bandwidth 3 kHz, 30 kHz, 300 kHz, and 5 MHz (–6 dB); 300 Hz with Opt. 01
RF Input 50 Ω, max. +20 dBm (0.1 W or 2.2 V) CW, ≤100 V DC (blocking capacitor)
RF Attenuator 0 dB to 50 dB, 2 dB steps
Sensitivity –117 dBm at 3 kHz RBW, –129 dBm at 3 kHz RBW w/preamp
Reference Level –70 dBm to +20 dBm; –23 dBmV to +66.9 dBmV
Sweep Speed 2 sec/Div to 1 µs/Div in 1–2–5 sequence
Triggering Modes Free Run, Internal, External, Line, TV Line, and TV Field
Displayed Average Noise −66 dBm to −100 dBm
Display Dynamic Range 80 dB
Dynamic Range ≥100 dBm
Internal Calibrator 100 MHz marker at –30 dBm
Weight Standard accessories: 11.25 kg (<25 lbs); Basic configuration: 9.5 kg (<21 lbs)
Power 90 − 132 VAC, 48 to 440 Hz; 90 – 250 VAC, 47 to 63 Hz, 90 W (standard) / 105 W max (with options)
Operating Environment 0 °C to +50 °C; Humidity: MIL-T-28800E; Altitude: Up to 4.6 km (15,000 ft)
Features
  • 28KB non-volatile memory (NVRAM) for up to 18 displays of waveforms and markers, and up to 9 front panel setups, user-defined key programs, and antenna factor tables
  • Single and delta markers for readout of frequency and amplitude
  • analog display capability, TV Line and TV Field triggering to provide demodulation of video carriers
  • internal audio amplifier and AM/FM detectors
  • digital storage (positive peak only, positive/negative peak, max hold, waterfall)
  • alternate reference level units (dBm, dBmV, dBV, dBµV, dBµW, and dBµV/m)
  • direct plot/print via Opt. 03 GPIB or Centronics interface Opt. 09

Options

  • Opt. 01: 300 Hz Resolution Bandwidth filter; Phase-lock with span/div ≤20 kHz
  • Opt. 02: Frequency Counter, 1 Hz or 1 kHz, menu selectable
  • Opt. 03: GPIB Interface (cannot be combined with Opt. 09)
  • Opt. 04: TG Tracking Generator, 100 kHz to 1.8 GHz (cannot be combined with Opt. 14)
  • Opt. 06: Battery Power Option (2704 Inverter and 2705 Battery Pack)
  • Opt. 08: Replace GPIB with RS-232-C (cannot be combined with Opt. 03)
  • Opt. 09: Centronix interface (cannot be combined with Opt. 03)
  • Opt. 10: Video Monitor mode
  • Opt. 11: Non-volatile RAM storage for setups, displays, and user-definable key sequences
  • Opt. 14: Adds Resolution Bandwidths of 1 kHz, 10 kHz, 100 kHz, and 1 MHz (cannot be combined with Opt. 04)
  • Opt. 15: Tek 1405 TV Sideband Adapter interface installed at J101 as SMA female
  • Opt. 30: Rackmount
  • Opt. 33: Travel Line Package
  • Opt. 34: Portable to Rackmount Adapter.

See also the 2706 preselector and 2707 tracking generator.

Connectors

Front panel: RF INPUT N connector (50 Ω, +20 dBm Max, 100 V DC blocking capacitor); GEN N connector (50 Ω, 0 to -48 dBm, Opt. 4 Tracking Generator generates a sweep frequency to enable the Spectrum Analyzer to make circuit response measurements, i.e. scalar network analysis).

Rear panel: J101 SMA connector (Female SMA installed for Opt. 15), J102 EXT TRIG BNC connector (10 kΩ, 35 V (DC + peak AC)), J103 DB9 connector (Accessory connector, not RS-232: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused), J104 GPIB connector

Links

Documents Referencing 2710

Document Class Title Authors Year Links
26W-7045.pdf Application Note Random Noise Measurement with the Spectrum Analyzer Morris Engelson 1989

Pictures