560-series scopes

From TekWiki
Revision as of 09:39, 22 April 2018 by Peter (talk | contribs)
Jump to navigation Jump to search

The Tektronix 560 series of mainframes includes the 560, 506, 561, 564, 565, 567, and 568. They were introduced with the 560 and 561 in 1961. The 560 series has plug-ins that drive the CRT deflection plates directly.

Early 560-series mainframes (e.g., the 561) had tubes in the low voltage power supply and in the high voltage power supply. Later scopes in the series (e.g., the 561B and 568) were all solid state except for the CRT.

Common Features

  • They accept 560-series plug-ins (the 567 additionally accepts the 6R1 or 6R1A plug-ins)
  • They require the plug-ins to drive the CRT vertical and horizontal deflection plates directly (exception: the 565, which includes built-in trigger, timebase, and horizontal amplifier circuitry)
  • They don't contain a vertical delay line (some plug-ins, e.g., the 3A6 have built-in delay lines so the rising edge of fast pulses with low repetition rate can be seen)
  • They have linear power supplies (exception: the 560 power supply is basically a switch-mode power supply)
  • They don't use post-deflection acceleration of the CRT beam

Plug-ins

560-series plug-ins includes the 2-series models (e.g., the 2A60 and 2B67) and the later 3-series models (e.g., the 3A1 and 3B3) which consume more power than the power supply in the 560 and 561 can deliver (the 561A can meet the current demands of 3-series plug-ins).

Advantages of plug-ins driving the CRT directly

  • Lower cost for an entry-level configuration. No need for expensive high-speed amplifiers anywhere in the system.
  • Bandwidth can be upgraded later by adding faster plug-ins.
  • Replacing plug-ins has a high likelihood of fixing problems, unless there is a problem in the power supply or CRT.
  • A small number of mainframe models can support a wide range of applications.

Disadvantages of plug-ins driving the CRT directly

  • Checking calibration after changing plug-ins is recommended.

The signal path of the plug-in to mainframe interface is not as well standardized as it is on 500-series scopes or 7000-series scopes. This is primarily due to variability in deflection sensitivity of CRTs. The 564B manual says:

The accuracy of measurements made with the Type 564B depends on the calibration of the plug-in units used. Since the plug-in units drive the deflection plated directly, each unit must be adjusted to match the deflection sensitivity of the particular CRT it drives. Therefore, the gain or sweep timing adjustment must be checked each time a plug-in unit is changed. On most units, gain or timing calibration is made with a front-panel screwdriver adjustment.

  • Since each plug-in contains deflection amplifiers, 560-series plug-ins cost more than a plug-in with equivalent functionality for a mainframe that has built-in deflection amplifiers. The purchase cost, weight, and maintenance burden of the extra deflection amplifiers adds up, particularly if the customer wants several different plug-ins at the high-speed end of the 560-series range.