5CT1N: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(Added 'Common Problems' Many 5CT1N and 7CT1N plugins suffer from this problem.)
No edit summary
Line 3: Line 3:
Common Problems
Common Problems


C30 and C32 (100 uF, 25V)have a tendency to leak or short, causing associated resistors R30 and R32 (240, .25W, 5%) to overheat and fail. A good preventative maintenance procedure would be to replace these two capacitors with modern equivalents. Two 5CT1N units purchased online had this fault, and replacement of the capacitors and resistors restored proper operation.
C30 and C32 (100 uF, 25V) have a tendency to leak or short, causing associated resistors R30 and R32 (240, .25W, 5%) to overheat and fail. A good preventative maintenance procedure would be to replace these two capacitors with modern equivalents. Two 5CT1N units purchased online had this fault, and replacement of the capacitors and resistors restored proper operation.

Revision as of 08:38, 2 September 2013

Curve Tracer Plug-in for the 5000 series oscilloscopes.

Common Problems

C30 and C32 (100 uF, 25V) have a tendency to leak or short, causing associated resistors R30 and R32 (240, .25W, 5%) to overheat and fail. A good preventative maintenance procedure would be to replace these two capacitors with modern equivalents. Two 5CT1N units purchased online had this fault, and replacement of the capacitors and resistors restored proper operation.