DC5004: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
(condensed specs)
 
(2 intermediate revisions by 2 users not shown)
Line 1: Line 1:
{{TM5000 | mfg=Tektronix | type=DC5004 | function=frequency counter | class=counter | image=dc5004.jpg | introduced=1989 | discontinued=(?) |
{{TM5000 | mfg=Tektronix | type=DC5004 | function=frequency counter | class=counter | image=dc5004.jpg | introduced=1989 | discontinued=(?) |
manuals=
designers= |manuals=
* Manual needed [[Category:Manual needed]] ''please add''
* Manual needed [[Category:Manual needed]] ''please add''
}} It is the TM5000 incarnation of the [[DC504A]].
}} It is the TM5000 incarnation of the [[DC504A]].


==Specification==
{{BeginSpecs}}
* 5 Digit LED Display
{{Spec | Frequency                     | DC (AC coupled: 10 Hz) to 100 MHz  }}
* Frequency Range
{{Spec | Sensitivity                   | 20 mV<sub>RMS</sub> below 15 MHz, 35 mV<sub>RMS</sub> below 50 MHz, down to 175 mV<sub>RMS</sub> at 80 MHz }}
** DC coupled: 0 Hz to at least 80 MHz
{{Spec | Input impedance               | 1 MΩ // 20 pF }}
** AC coupled: 10 Hz to at least 80 MHz
{{Spec | Trigger Level                 | ±1.5 V }}
* Sensitivity (sine wave):
{{Spec | Frequency Resolution         | 5 digits (LED Display); kHz ranges: 0.1 Hz; 1 Hz and 10 Hz (7 digit resolution possible in overflow kHz position); MHz ranges: 100 Hz, 1 kHz (9 digit resolution possible in overflow kHz position) }}
** 20 mV RMS (56,6 mV_pp) below 15 MHz
{{Spec | Display Time        | Variable  ~0.1 s ... ~10 s and HOLD mode}}
** 35 mV RMS (99 mV_pp) at or below 50 MHz
{{Spec | Stability (standard timebase) | 10<sup>–5</sup> (0°C to +50°C after 30 min warmup), drift <10<sup>–5</sup> per month }}
** 35 mV RMS @ 50 MHz derated down to 175 mV RMS (495 mV_pp) @80 MHz
{{Spec | Accuracy (standard timebase) | Adjustable to within 10<sup>–7</sup> }}
* Input impedance: 1 MΩ; 20 pF
{{Spec | Stability (Opt. 1)            | 10<sup>–7</sup> (0°C to +50°C after 30 min warmup), drift <10<sup>–7</sup> per month }}
* Trigger Level Range: -1,5V to +1,5 V
{{Spec | Accuracy (Opt. 1)            | Adjustable to within 5×10<sup>–9</sup> }}
*Maximum Resolution
{{EndSpecs}}
**Frequency kHz Positions: 0,1 Hz; 1 Hz and 10 Hz (Seven digit resolution possible in overflow kHz position)
**Frequency MHz Positions: 100 Hz and 1 kHz (Nine digit resolution possible in overflow kHz position)
**Period msec Positions: 1µs and 10 µs
**Period sec Positions: 10 ms; 1ms and 0,1 ms
*Trigger Error: 0,5% of one cycle at min, trigger sensitivity
*Internal Standard Time Base (1 MHz)
** Stability: 0°C to +50°C after 30 min warm up: 1E-5
** Long Term Drift: 1E-5 per month
** Accuracy: Adjustable to within 1E-7
*Internal  Time Base Option 1 (5 MHz)
** Stability: 0°C to +50°C after 30 min warm up: eE-7
** Long Term Drift: 1E-7 per month
** Accuracy: Adjustable to within 5E-9
* Display Time: Variable  ~0,1 sec ... ~10 sec and HOLD mode


==Pictures==
==Pictures==
Line 35: Line 21:
Dc5004 versus dc504a.png|Description of DC5004 from 1989 Catalog
Dc5004 versus dc504a.png|Description of DC5004 from 1989 Catalog
</gallery>
</gallery>
==Components==
{{Parts|DC5004}}

Latest revision as of 08:39, 6 December 2023

Tektronix DC5004
frequency counter
Tektronix DC5004

Produced from 1989 to (?)

Manuals
  • Manual needed please add
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix DC5004 is a frequency counter plug-in for the TM5000 system. It is the TM5000 incarnation of the DC504A.

Key Specifications

Frequency DC (AC coupled: 10 Hz) to 100 MHz
Sensitivity 20 mVRMS below 15 MHz, 35 mVRMS below 50 MHz, down to 175 mVRMS at 80 MHz
Input impedance 1 MΩ // 20 pF
Trigger Level ±1.5 V
Frequency Resolution 5 digits (LED Display); kHz ranges: 0.1 Hz; 1 Hz and 10 Hz (7 digit resolution possible in overflow kHz position); MHz ranges: 100 Hz, 1 kHz (9 digit resolution possible in overflow kHz position)
Display Time Variable ~0.1 s ... ~10 s and HOLD mode
Stability (standard timebase) 10–5 (0°C to +50°C after 30 min warmup), drift <10–5 per month
Accuracy (standard timebase) Adjustable to within 10–7
Stability (Opt. 1) 10–7 (0°C to +50°C after 30 min warmup), drift <10–7 per month
Accuracy (Opt. 1) Adjustable to within 5×10–9

Pictures

Components

Some Parts Used in the DC5004

Part Part Number(s) Class Description Used in
Intersil ICM7226 Monolithic integrated circuit 8-decade counter and 7-segment driver DC504A DC5004