{{Spec | AC V accuracy | 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C }}
{{Spec | AC V accuracy | 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C }}
{{Spec | AC V input resistance | 10 MΩ // 160 pF }}
{{Spec | AC V input resistance | 10 MΩ // 160 pF }}
{{Spec | dB ranges | -40, -20, 0, +20, +40 dB }}
{{Spec | dB ranges (DM501A only) | -40, -20, 0, +20, +40 dB }}
{{Spec | Ω ranges | 200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20  MΩ }}
{{Spec | Ω ranges | 200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20  MΩ }}
{{Spec | Ω accuracy | 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C }}
{{Spec | Ω accuracy | 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C }}
Line 33: Line 33:
Tek dm501a on.jpg|DM501A
Tek dm501a on.jpg|DM501A
Tek dm501a 1.jpg | DM501A
Tek dm501a 1.jpg | DM501A
Tek dm501a ext1.jpg|DM501A on an extension cable
Tek dm501a dark.jpg|DM501
</gallery>
</gallery>


* [http://w140.com/tekwiki/images/d/d5/Siliconix_LD120_and_LD121A_datasheets.pdf Siliconix LD120 and LD121A datasheets]
* [http://w140.com/tekwiki/images/d/d5/Siliconix_LD120_and_LD121A_datasheets.pdf Siliconix LD120 and LD121A datasheets]

Revision as of 10:47, 17 April 2018

Tektronix DM501
4½ digit True RMS multimeter
Tektronix DM501

Produced from 1978 to 1984

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix DM501 is a 4½ digit True RMS multimeter plug-in for the TM500 system.


It also has a temperature input, requiring the P6058 or P6430 temperature probe (DM501A: P6601 PT-100 probe).

Key Specifications

DC V ranges 200 mV, 2 V, 20 V, 200 V, 1000 V
DC V accuracy 0.05% of reading + 0.01% FS in 2 V to 200 V ranges @ +18 °C to +28 °C
DC V input resistance 10 MΩ
AC V ranges 200 mV, 2 V, 20 V, 200 V, 500 V
AC V accuracy 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C
AC V input resistance 10 MΩ // 160 pF
dB ranges (DM501A only) -40, -20, 0, +20, +40 dB
 Ω ranges 200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20  MΩ
 Ω accuracy 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C
DC A ranges 200 μA, 2 mA, 20 mA, 200 mA, 2 A
DC A accuracy 0.2% of reading + 0.015% FS @ +18 °C to +28 °C
AC A ranges 200 μA, 2 mA, 20 mA, 200 mA, 2 A
AC A accuracy 0.6% of reading + 0.05% FS, 20 Hz to 10 kHz @ +18 °C to +28 °C
Temperature range -62 °C to +240 °C
Temperature accuracy ±2 °C from -62 °C to +150 °C with probe calibrated to instrument, @ +18 °C to +28 °C ambient