==R10 ([[307-1015-00]] input voltage divider) Repair==
Fault pattern:  V<sub>DC</sub>, V<sub>AC</sub> and resistance (exept the lowest range) are 2% to low.
Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance.
Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath.


==See Also==
==See Also==
Line 12: Line 19:
Tek-1260-1971-00-rep4.jpg| Defect push button plunger on top and new one below
Tek-1260-1971-00-rep4.jpg| Defect push button plunger on top and new one below
</gallery>
</gallery>
[[Category:Instrument repair reports]]

Latest revision as of 10:14, 4 September 2023

Switch Repair

The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches. Refer to the the pictures and the repair section of switch assembly 260-1971-00.

R10 (307-1015-00 input voltage divider) Repair

Fault pattern:  VDC, VAC and resistance (exept the lowest range) are 2% to low.

Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance.

Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath.