P6008: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
mNo edit summary
Line 23: Line 23:
{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Attenuation | 10 × ±3% with 1 MΩ ±2% scope input }}
{{Spec | Attenuation | 10 × ±3% with 1 MΩ ±2% scope input }}
{{Spec | Maximum Voltage | 600 V DC or AC p-p, derated above 20 MHz }}
{{Spec | Compensation range | 15 pF (may be compensated to 20 pF with some loss of bandwidth) }}
{{Spec | Compensation range | 15 pF (may be compensated to 20 pF with some loss of bandwidth) }}
{{Spec | Bandwidth | 66 MHz w/Type 82 or 86 Plugin (63 MHz with x10 GAIN) }}
{{Spec | Bandwidth | 66 MHz w/Type 82 or 86 Plugin (63 MHz with x10 GAIN) }}
{{Spec | Input capacitance | ~7 pF }}  
{{Spec | Input capacitance | ~7 pF }}
{{EndSpecs}}
{{EndSpecs}}


Line 35: Line 36:
{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Attenuation | 10 × ±3% with 1 MΩ ±2% scope input }}
{{Spec | Attenuation | 10 × ±3% with 1 MΩ ±2% scope input }}
{{Spec | Maximum Voltage | 600 V, DC + peak AC, derated with frequency }}
{{Spec | Compensation range | 15 pF to 55 pF }}
{{Spec | Compensation range | 15 pF to 55 pF }}
{{Spec | Bandwidth | 100 MHz w/15 pF scope inputs }}
{{Spec | Bandwidth | 100 MHz w/15 pF scope inputs }}
{{Spec | Input capacitance | 10-14 pF, depending on scope input capacitance  }}  
{{Spec | Input capacitance | 10-14 pF, depending on scope input capacitance  }}
{{EndSpecs}}  
{{EndSpecs}}


==Pictures==
==Pictures==

Revision as of 14:13, 7 December 2021

Tektronix P6008
Passive 66 MHz, x10 probe
Early P6008

Compatible with 1 MΩ scope inputs

Produced from (?) to (?)

Manuals


Manuals – Specifications – Links – Pictures

There are two different versions, or generations, of the Tektronix P6008 ×10 passive voltage probe. Each was fitted with a BNC connector with no readout encoding pin.

010-0130-00

The early (1963) P6008 Probe (010-130 & 010-0130-00) was primarily designed for use with the Type 82 and Type 86 plug-ins in a 580-series scope, with 15 pF inputs. It came with a 3.5 foot gray cable. Like the other members of the monolithic probe family, the early P6008 was rated for temperatures up to 75 C.

Key Specifications

Attenuation 10 × ±3% with 1 MΩ ±2% scope input
Maximum Voltage 600 V DC or AC p-p, derated above 20 MHz
Compensation range 15 pF (may be compensated to 20 pF with some loss of bandwidth)
Bandwidth 66 MHz w/Type 82 or 86 Plugin (63 MHz with x10 GAIN)
Input capacitance ~7 pF

010-0130-01

The later (1971) P6008 ENVIRONMENTAL PROBE (010-0130-01) can be compensated to a wider range of scope input capacitances, has wider bandwidth, but higher input capacitance. It came with a 6 foot white cable. This later version of the P6008 was rated for temperatures up to 150 C for the probe and cable and up to 55 C for the compensation box.

Key Specifications

Attenuation 10 × ±3% with 1 MΩ ±2% scope input
Maximum Voltage 600 V, DC + peak AC, derated with frequency
Compensation range 15 pF to 55 pF
Bandwidth 100 MHz w/15 pF scope inputs
Input capacitance 10-14 pF, depending on scope input capacitance