P6016: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
m (Added use with a Type 134)
Line 14: Line 14:
[[File:Tek 131 front.jpg|thumb|250px|right|Type 131 Current Probe Amplifier for P6016]]
[[File:Tek 131 front.jpg|thumb|250px|right|Type 131 Current Probe Amplifier for P6016]]


The '''Tektronix P6016''' is a 125-turn AC current probe for oscilloscopes. It must be used with either the 011-028 passive termination or the [[131|Type 131 Current Probe Amplifier]].
The '''Tektronix P6016''' is a 125-turn AC current probe for oscilloscopes. It must be used with either the 011-028 passive termination or the [[131|Type 131 Current Probe Amplifier]]. Brief testing suggests that the P6016 works quite well, and possibly bettter, with the Type [[134]] in P6019/P6021 mode.


{{BeginSpecs}}
{{BeginSpecs}}

Revision as of 13:28, 31 March 2023

Tektronix P6016
Oscilloscope Current Probe
P6016 Probe

Compatible with 131

Produced from 1960 to (?)

Manuals


Manuals – Specifications – Links – Pictures
011-028 Passive Termination for P6016
Type 131 Current Probe Amplifier for P6016

The Tektronix P6016 is a 125-turn AC current probe for oscilloscopes. It must be used with either the 011-028 passive termination or the Type 131 Current Probe Amplifier. Brief testing suggests that the P6016 works quite well, and possibly bettter, with the Type 134 in P6019/P6021 mode.

Key Specifications

Maximum current 15 Ap-p
DC saturation 0.5 A
     — With the 011-028 Passive Termination
Sensitivity 2 mA/mV or 10 mA/mV selectable
Rise time 18 ns (eq. 20 MHz -3 dB)
Low frequency -3 dB point Switchable between 850 Hz at 2 mA/mV or 230 Hz at 10 mA/mV
Delay time 12 ns typ.
     — With a Type 131 Current Probe Amplifier
Sensitivity 1 mA/Div to 1 A/Div, 1-2-5
Noise 0.1 mAp-p
Rise time 20 ns (eq. 17 MHz -3 dB)
Low frequency -3 dB point 50 Hz
Delay time 32 ns typ.

Pictures