P6021: Difference between revisions

From TekWiki
Jump to navigation Jump to search
mNo edit summary
 
(15 intermediate revisions by 2 users not shown)
Line 1: Line 1:
[[File:Tek-p6021.jpg|thumb|350px|right|P6021 with passive termination box]]
{{Probe Sidebar
The '''Tektronix P6021''' is a 125-turn AC current probe for oscilloscopes.
|manufacturer=Tektronix
It can be used with the included passive termination box (011-0105-00),
|model=P6021
a [[134|type 134 current probe amplifier]], or a [[3A9]] or [[7A14]] current probe amplifier scope plug-in.  
|summary=oscilloscope current probe
|image=Tek-p6021.jpg
|caption=P6021 with Passive Termination Box
|for=134
|introduced=1969
|discontinued=(?)
|manuals=
* [[Media:070-0947-00.pdf|Tektronix P6021 Manual -00]] (1981 version)
* [[Media:070-0947-05.pdf|Tektronix P6021 Manual -05]] (earlier version with termination schematics)
* [[Media:070-0947-06.pdf|Tektronix P6021 Manual -06]] (no schematics)
* [[Media:070-0947-07.pdf|Tektronix P6021 Manual -07]] (no schematics)
* [[Media:Tek_p6021_1971_cat.pdf|P6021 Description in 1971 Catalog]]
}}
The '''Tektronix P6021''' is a 125-turn AC current probe for oscilloscopes. It can be used with the included passive termination box (011-0105-00), a [[134|Type 134 Current Probe Amplifier]], or a [[3A9]] or [[7A14]] current probe amplifier scope plug-in.  


Bandwidth is up to 52 MHz.
Bandwidth is up to 52 MHz.
Line 13: Line 26:
{| class="wikitable"
{| class="wikitable"
!  
!  
! P6021 passive
! P6021 + 011-0105-00 passive termination
! P6021 + [[134]]
! P6021 + [[134]]
! P6021 + [[7A14]]
! P6021 + [[7A14]]
! P6022 passive
! [[P6022]] + [[011-0106-00]] passive termination
! P6022 + [[134]]
! [[P6022]] + [[134]]
! P6022 + [[7A14]]
! [[P6022]] + [[7A14]]
|-
|-
! Sensitivity
! Sensitivity
Line 24: Line 37:
| 1 mA/mV or 10 mA/mV  || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div)  || 1 mA/Div to 1 A/Div
| 1 mA/mV or 10 mA/mV  || 1 mA/Div to 1 A/Div<br/>(scope @ 50 mV/Div)  || 1 mA/Div to 1 A/Div
|-
|-
! Bandwidth (-3 dB)
! Bandwidth (–3 dB)
| 60 MHz @ 10 mA/mV    ||  > 40 MHz  ||  50 MHz
| 60 MHz @ 10 mA/mV    ||  > 40 MHz  ||  50 MHz
| 200 MHz @ 10 mA/mV  ||  > 70 MHz  ||  105 MHz
| 200 MHz @ 10 mA/mV  ||  > 70 MHz  ||  105 MHz
|-
|-
! LF limit (-3 dB)
! LF limit (–3 dB)
| 450 Hz @ 2 mA/mV<br/>120 Hz @ 10 mA/mV  ||  < 12 Hz  ||  25 Hz
| 450 Hz @ 2 mA/mV<br/>120 Hz @ 10 mA/mV  ||  < 12 Hz  ||  25 Hz
| 8500 Hz @ 1 mA/mV<br/>935 Hz @ 10 mA/mV  ||  < 100 Hz  ||  250 Hz
| 8500 Hz @ 1 mA/mV<br/>935 Hz @ 10 mA/mV  ||  < 100 Hz  ||  250 Hz
Line 37: Line 50:
|}
|}


==Manuals==
==Links==
 
{{Documents|Link=P6021}}
* [[Media:070-0947-00.pdf|Tektronix P6021 Manual (1981 version, PDF)]]
* [[Media:TEK P6021 Manual.pdf | Tektronix P6021 Manual (PDF), earlier version with termination schematics]]
* [http://w140.com/tek_P6021.pdf Tektronix P6021 Manual (PDF), no schematics]
* [http://w140.com/tek_p6021_1971_cat.pdf Description in 1971 Catalog (PDF)]


==Prices==
==Prices==
Line 50: Line 59:
! 1970
! 1970
! 1974
! 1974
! 1984
! 1986
! 1986
! 1988
|-
|-
! Catalog price
! Catalog price
|width=4em align=right|  $121
|width=4em align=right|  $121
|width=4em align=right|  $100
|width=4em align=right|  $100
|width=4em align=right|  $345
|width=4em align=right|  $430
|width=4em align=right|  $430
|width=4em align=right|  $450
|-
|-
! 2017 value
! 2021 value
|align=right|  $760
|align=right|  $850
|align=right|  $500
|align=right|  $550
|align=right|  $960
|align=right|  $910
|align=right| $1,070
|align=right| $1,040
|-
|-
|}
|}
Line 74: Line 89:
<gallery>
<gallery>
P6021-bat41-1.jpg | Measuring diode reverse recovery
P6021-bat41-1.jpg | Measuring diode reverse recovery
P6021-bat41-2.jpg | BAT41 diode reverse recovery, 20 mA/Div
P6021-bat41-2.jpg | Small-signal diode reverse recovery, 20 mA/Div
</gallery>
</gallery>
Comparison of P6022 with passive termination @ 10 mA/Div (top trace) vs. P6021+134 (bottom trace).<br />
Comparison of P6022 with passive termination @ 10 mA/Div (top trace) vs. P6021+134 (bottom trace).<br />
The signal shown is a sawtooth from a HP3325A @ 5 V<sub>p-p</sub> into 50 Ω = 100 mA<sub>p-p</sub>.  
The signal shown is a sawtooth from a HP3325A @ 5 V<sub>p-p</sub> into 50 Ω = 100 mA<sub>p-p</sub>.  
<gallery>
<gallery>
P6021-6022-100hz.jpg | 100 Hz
P6021-6022-100hz.jpg   | 100 Hz
P6021-6022-1000hz.jpg | 1 kHz
P6021-6022-1000hz.jpg | 1 kHz
P6021-6022-10000hz.jpg | 10 kHz
P6021-6022-10000hz.jpg | 10 kHz
</gallery>
</gallery>


[[Category:Oscilloscope current probes]]
[[Category:Oscilloscope current probes]]
[[Category:Introduced in 1969]]

Latest revision as of 22:10, 2 December 2023

Tektronix P6021
oscilloscope current probe
P6021 with Passive Termination Box

Compatible with 134

Produced from 1969 to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6021 is a 125-turn AC current probe for oscilloscopes. It can be used with the included passive termination box (011-0105-00), a Type 134 Current Probe Amplifier, or a 3A9 or 7A14 current probe amplifier scope plug-in.

Bandwidth is up to 52 MHz.

The P6021 has a contact to short the ring around the 7A14's BNC input to ground, to switch frequency compensation and gain.

The sister model P6022 has higher bandwidth but also a higher LF cut-off frequency and lower peak current rating.

Specifications

P6021 + 011-0105-00 passive termination P6021 + 134 P6021 + 7A14 P6022 + 011-0106-00 passive termination P6022 + 134 P6022 + 7A14
Sensitivity 2 mA/mV or 10 mA/mV 1 mA/Div to 1 A/Div
(scope @ 50 mV/Div)
1 mA/Div to 1 A/Div 1 mA/mV or 10 mA/mV 1 mA/Div to 1 A/Div
(scope @ 50 mV/Div)
1 mA/Div to 1 A/Div
Bandwidth (–3 dB) 60 MHz @ 10 mA/mV > 40 MHz 50 MHz 200 MHz @ 10 mA/mV > 70 MHz 105 MHz
LF limit (–3 dB) 450 Hz @ 2 mA/mV
120 Hz @ 10 mA/mV
< 12 Hz 25 Hz 8500 Hz @ 1 mA/mV
935 Hz @ 10 mA/mV
< 100 Hz 250 Hz
Maximum current 15 Ap-p 6 Ap-p
(> 10 kHz @ 1 mV/mA, > 3 kHz @ 10 mV/mA, < 10 MHz)

Links

Documents Referencing P6021

Prices

Year 1970 1974 1984 1986 1988
Catalog price $121 $100 $345 $430 $450
2021 value $850 $550 $910 $1,070 $1,040

Pictures

Measurement examples

Comparison of P6022 with passive termination @ 10 mA/Div (top trace) vs. P6021+134 (bottom trace).
The signal shown is a sawtooth from a HP3325A @ 5 Vp-p into 50 Ω = 100 mAp-p.