The S-3 sampling head is a plug-in head for use with the  
The '''Tektronix S-3 sampling head''' is a plug-in head for use with the  
7000 series sampling plug-ins or with the Type [[3S2]], [[3S5]], or [[3S6]] samplers.  
7000 series sampling plug-ins or the Type [[3S2]], [[3S5]], and [[3S6]] samplers.  
The plug-in includes a permanently attached 1× probe with the sampling bridge enclosed in the probe. Optional 10× or 100× attenuator tips are available.
 
The plug-in includes a permanently attached 1× probe with the sampling bridge enclosed in the probe.
Optional 10× or 100× attenuator tips (010-0364-00, 010-0365-00) provide extended input voltage range.


==Specifications==
==Specifications==
Line 27: Line 29:
* Maximum signal voltage: 2 V<sub>p-p</sub>
* Maximum signal voltage: 2 V<sub>p-p</sub>
* The S-3 does not provide a trigger pickoff for internal triggering.
* The S-3 does not provide a trigger pickoff for internal triggering.
* An offset voltage of either +/- 1 V or +/- 2 V can be selected via a toggle switch on the front panel.
* An offset voltage of either ±1 V or ±2 V can be selected via a toggle switch on the front panel.


== Differences between S-3 and S-3A ==
== Differences between S-3 and S-3A ==
The S-3A sampling head has a 10&times; improvement in baseline shift due to cable flexing. All other specifications are identical to the ordinary S-3. The major differences in design involves the gate board in the box part of the probe, but some changes are also made to the circuits in the body of the probe.
The S-3A sampling head has a 10&times; improvement in baseline shift due to cable flexing.
All other specifications are identical to the ordinary S-3.
 
The major differences in design involves the gate board in the box part of the probe,
but some changes are also made to the circuits in the body of the probe.


== Difference between the S-3 and the P6038 ==
== Difference between the S-3 and the P6038 ==
The S-3 probe contains a compensated attenuator before the sampling bridge. The [[P6038]] does not.   
The S-3 probe contains a compensated attenuator before the sampling bridge.
The S-3 probe has a JFET buffer between the holding cap and the probe cable. The P6038 does not.
The [[P6038]] does not.   
The S-3 probe has a JFET buffer between the holding cap and the probe cable.
The P6038 does not.
Both use the [[152-0368-00]] sampling bridge.
Both use the [[152-0368-00]] sampling bridge.


Line 42: Line 50:
==Pictures==
==Pictures==
<gallery>
<gallery>
Image:S3_frontend.png
S3_frontend.png
File:S3a attenuator.jpg|S-3A attenuator
S3a attenuator.jpg|010-0364-00 ×10 attenuator for S-3A
File:S3a box front.jpg|Front
S3a box front.jpg|Front
File:S3a probe external1.jpg|S-3A probe
S3a probe external1.jpg|S-3A probe
File:S3a probe internal1.jpg|S-3A probe internal
S3a probe internal1.jpg|S-3A probe internal
File:S3a probe internal2.jpg|S-3A probe internal
S3a probe internal2.jpg|S-3A probe internal
File:S3a strobe board.jpg|S-3A strobe board
S3a strobe board.jpg|S-3A strobe board
File:S3a probe_internal_bothsides.jpg|S-3A probe both sides
S3a probe_internal_bothsides.jpg|S-3A probe both sides
File:Tek s3 probe sample bridge.jpg|S-3 probe sampling bridge
Tek s3 probe sample bridge.jpg|S-3 probe sampling bridge
File:Tek s3 sample bridge2.jpg|S-3 probe sampling bridge
Tek s3 sample bridge2.jpg|S-3 probe sampling bridge
</gallery>
</gallery>


[[Category:7000 and 3S series sampling heads]]
[[Category:7000 and 3S series sampling heads]]

Revision as of 04:50, 28 February 2018

Template:Plugin Sidebar 2 The Tektronix S-3 sampling head is a plug-in head for use with the 7000 series sampling plug-ins or the Type 3S2, 3S5, and 3S6 samplers.

The plug-in includes a permanently attached 1× probe with the sampling bridge enclosed in the probe. Optional 10× or 100× attenuator tips (010-0364-00, 010-0365-00) provide extended input voltage range.

Specifications

  • Input rise time:
    • 1×: < 350 ps
    • 10×: < 400 ps
    • 100×: < 500 ps
  • Input impedance:
    • 1×: 100 kΩ || 2.3 pF at low frequency
    • 10×: 1 MΩ || 2 pF
    • 100×: 1 MΩ || 1.7 pF
  • Maximum signal voltage: 2 Vp-p
  • The S-3 does not provide a trigger pickoff for internal triggering.
  • An offset voltage of either ±1 V or ±2 V can be selected via a toggle switch on the front panel.

Differences between S-3 and S-3A

The S-3A sampling head has a 10× improvement in baseline shift due to cable flexing. All other specifications are identical to the ordinary S-3.

The major differences in design involves the gate board in the box part of the probe, but some changes are also made to the circuits in the body of the probe.

Difference between the S-3 and the P6038

The S-3 probe contains a compensated attenuator before the sampling bridge. The P6038 does not. The S-3 probe has a JFET buffer between the holding cap and the probe cable. The P6038 does not. Both use the 152-0368-00 sampling bridge.

Repair of Sampling Diodes

Individuals have reported success in replacing the S-3 sampling diodes with Agilent matched pair Schottky diodes.