067-0500-00/Repairs: Revision history

Jump to navigation Jump to search

Diff selection: Mark the radio buttons of the revisions to compare and hit enter or the button at the bottom.
Legend: (cur) = difference with latest revision, (prev) = difference with preceding revision, m = minor edit.

27 September 2023

26 September 2023

  • curprev 17:4617:46, 26 September 2023Jadney talk contribs 946 bytes +946 Created page with "In order to allow my 3M1 to work with 40 V pulses, I added a second resistor in parallel to R104, increasing the quiescent bias point of the 6DJ8. The resistor I added was a 620 kΩ, and the resulting parallel combination measures ~185 kΩ. This now works perfectly in my 564B. It should work equally well in any 560 series scope. You can see the extra resistor piggy-backed on R104 in the last two photos of the 3M1. It's worth noting that the input circuit in the 3M1 a..."