TWD120: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(Created page with "The Tektronix TWD 120 Digitizer allows a Windows PC to be used as a digital storage oscilloscope. It is rated at 100 MHz, has two input channels and an external trigger input...")
 
(Added firmware update info)
 
(16 intermediate revisions by 4 users not shown)
Line 1: Line 1:
The Tektronix TWD 120 Digitizer allows a Windows PC to be used as a digital storage oscilloscope. It is rated at 100 MHz, has two input channels and an external trigger input. It connects to the host PC via a SCSI bus.  
{{Oscilloscope Sidebar
|manufacturer=Tektronix
|series=
|model=TWD120
|summary=100 MHz, 100 MS/s 8-bit 2-ch digitizer 
|image=twd120 9.jpg
|caption=Tektronix TWD120
|introduced=1994
|discontinued=1995
|designers=
|manuals=
* [http://www.ko4bb.com/getsimple/index.php?id=download&file=Tektronix/Tektronix_Miscelaneous/Tektronix_TWD120_Digitizer_Operator_Manual.zip Tektronix TWD 120 Digitizer Instruction Manual (ZIP)]
* [http://www.ko4bb.com/getsimple/index.php?id=download&file=Tektronix/Tektronix_TWD120_Digitizer_Operator_Manual.pdf Tektronix TWD 120 Digitizer Reference Manual (PDF)]
}}
The '''Tektronix TWD 120 Digitizer''' allows a Windows PC to be used as a digital storage oscilloscope. It is rated at 100 MHz, has two input channels and an external trigger input. It connects to the host PC via a SCSI bus.  


The TWD 120 is about the size of a laptop, and was supplied with a soft case. It was intended to be used in conjunction with a laptop running Windows 3.1, and was apparently manufactured for GE for use by medical device field technicians.  
The TWD 120 is about the size of a laptop, and was supplied with a soft case. It was intended to be used in conjunction with a laptop running Windows 3.1, and was apparently manufactured for GE for use by medical device field technicians.  
Line 5: Line 19:
The Adaptec ASPI SCSI driver is required for the software to communicate with the digitizer. SCSI configuration can be tricky.
The Adaptec ASPI SCSI driver is required for the software to communicate with the digitizer. SCSI configuration can be tricky.


Firmware is updated using the [[067-1436-00|067-1436-00 Communications Pod]]
{{BeginSpecs}}
{{SpecGroup|Acquisition}}
{{Spec | Bandwidth | 100 MHz (20 MHz selectable limit) }}
{{Spec | Sample rate | 2 channels @ 100 MS/s each (two digitizers) }}
{{Spec | Resolution | 8 bit }}
{{Spec | Record length | 500 to 5000 samples per channel }}
{{Spec | Modes | sample, envelope, average (16 samples), high-resolution, and peak-detect }}
{{SpecGroup | Inputs}}
{{Spec | Deflection | 2 mV/div - 99.5 mV/div (±1 V offset), 100 mV/div - 995 mV/div (±10 V offset), 1 V/div - 10 V/div (±100 V offset) }}
{{Spec | Max. input | ±400&thinsp;V (DC + peak AC) up to 100&thinsp;kHz, derate −20&thinsp;dB/decade up to 3&thinsp;MHz, 13&thinsp;V<sub>p-p</sub> above 3&thinsp;MHz }}
{{Spec | Input impedance | 1&thinsp;MΩ }}
{{SpecGroup | Trigger}}
{{Spec | Trigger modes | edge, error, external and video (NTSC, SECAM, PAL, Sync negative or positive) }}
{{Spec | Trigger Source | CH1, CH2, External }}
{{Spec | Trigger coupling |  AC, noise reject, high frequency reject, low frequency reject }}
{{Spec | Trigger levels | ±12 divisions (CH1, CH2) or up to ±15&thinsp;V (External) }}
{{Spec | Trigger holdoff |  500&thinsp;ns to 10&thinsp;s }}
{{SpecGroup | Misc }}
{{Spec | Interfaces |
* [[Ultra SCSI II]] (50-pin high density [[SCSI connector]]) for host connection
* [[RS-232 interface]] (DE-9)
* [[Parallel printer interface]] (DB-25)
}}
{{Spec | Calibrator output | 5&thinsp;V @ 1&thinsp;kHz (1&thinsp;MΩ load) }}
{{Spec | Power | 12−24&thinsp;V<sub>DC</sub>, 30&thinsp;W max }}
{{EndSpecs}}


== Links ==
== Links ==
http://www.chichak.ca/file/Tek_TWD120.html - good information on SCSI configuration and work arounds for using later versions of Windows
* [http://www.chichak.ca/file/Tek_TWD120.html More info, including good information on SCSI configuration and work arounds for using later versions of Windows]
https://www.eevblog.com/forum/reviews/tektronix-twd120-100mhz-dual-channel-digitizer-teardown-opinions/ - teardown comments and pictures
* [https://www.eevblog.com/forum/reviews/tektronix-twd120-100mhz-dual-channel-digitizer-teardown-opinions/ EEVblog teardown comments and pictures]
 
==Photos==
<gallery>
Tek twd120 1.jpg
Tek twd120 2.jpg
Tek twd120 3.jpg
Tek twd120 4.jpg
Tek twd120 5.jpg
Tek twd120 6.jpg
Tek twd120 7.jpg
Tek twd120 8.jpg
twd120 9.jpg
</gallery>
 
 
[[Category:Digital storage scopes]]

Latest revision as of 06:08, 5 October 2023

Tektronix TWD120
100 MHz, 100 MS/s 8-bit 2-ch digitizer
Tektronix TWD120

Produced from 1994 to 1995

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix TWD 120 Digitizer allows a Windows PC to be used as a digital storage oscilloscope. It is rated at 100 MHz, has two input channels and an external trigger input. It connects to the host PC via a SCSI bus.

The TWD 120 is about the size of a laptop, and was supplied with a soft case. It was intended to be used in conjunction with a laptop running Windows 3.1, and was apparently manufactured for GE for use by medical device field technicians.

The Adaptec ASPI SCSI driver is required for the software to communicate with the digitizer. SCSI configuration can be tricky.

Firmware is updated using the 067-1436-00 Communications Pod

Key Specifications

     — Acquisition
Bandwidth 100 MHz (20 MHz selectable limit)
Sample rate 2 channels @ 100 MS/s each (two digitizers)
Resolution 8 bit
Record length 500 to 5000 samples per channel
Modes sample, envelope, average (16 samples), high-resolution, and peak-detect
     — Inputs
Deflection 2 mV/div - 99.5 mV/div (±1 V offset), 100 mV/div - 995 mV/div (±10 V offset), 1 V/div - 10 V/div (±100 V offset)
Max. input ±400 V (DC + peak AC) up to 100 kHz, derate −20 dB/decade up to 3 MHz, 13 Vp-p above 3 MHz
Input impedance 1 MΩ
     — Trigger
Trigger modes edge, error, external and video (NTSC, SECAM, PAL, Sync negative or positive)
Trigger Source CH1, CH2, External
Trigger coupling AC, noise reject, high frequency reject, low frequency reject
Trigger levels ±12 divisions (CH1, CH2) or up to ±15 V (External)
Trigger holdoff 500 ns to 10 s
     — Misc
Interfaces
Calibrator output 5 V @ 1 kHz (1 MΩ load)
Power 12−24 VDC, 30 W max

Links

Photos