Thermal cutoff: Difference between revisions
No edit summary |
No edit summary |
||
(8 intermediate revisions by 3 users not shown) | |||
Line 1: | Line 1: | ||
Thermal cutoff switches are common in all Tektronix oscilloscopes | [[File:Thermal cutout switch.jpg|thumb|450px|right|Thermal cutout switch in an [[RM502A]] scope]] | ||
starting in the mid- | '''Thermal cutoff switches''' are common in all Tektronix oscilloscopes | ||
temperature, typically about | starting in the mid-1950s. If the chassis gets above a certain | ||
temperature, typically about 53 − 56 °C (128 − 133°F), the switch | |||
opens. Typically this interrupts the mains power coming into the | opens. Typically this interrupts the mains power coming into the | ||
scope and the scope will cool down. Thermal cutoff switches used in | scope and the scope will cool down. | ||
Thermal cutoff switches used in | |||
Tek scopes are self-resetting. After the scope cools down sufficiently, | Tek scopes are self-resetting. After the scope cools down sufficiently, | ||
the thermal switch will close, allowing power into the instrument. | the thermal switch will close, allowing power into the instrument. | ||
Instruments without thermal cutoff switches are the exception. They | Instruments without thermal cutoff switches are the exception. | ||
are the [[511]], [[512]], [[513]], | They are the [[511]], [[512]], [[513]], [[514]], [[517]] (serial numbers 1739 and below), | ||
and possibly others. | |||
Certainly all of the 500-series scopes that take | Certainly all of the 500-series scopes that take | ||
[[letter-series and 1-series plug-ins]] have thermal cutoff switches. | [[letter-series and 1-series plug-ins]] have thermal cutoff switches. | ||
This also includes the [[127]], [[132]], and [[133]] powered plug-in enclosures. | |||
The [[310|310A]] has an unusually high 165°F (74°C) thermal cutoff. | |||
This is due to the fact that the 310A is a compact tube scope without a fan. | |||
[[Category:Electromechanical components]] |
Latest revision as of 12:33, 1 September 2024
Thermal cutoff switches are common in all Tektronix oscilloscopes starting in the mid-1950s. If the chassis gets above a certain temperature, typically about 53 − 56 °C (128 − 133°F), the switch opens. Typically this interrupts the mains power coming into the scope and the scope will cool down.
Thermal cutoff switches used in Tek scopes are self-resetting. After the scope cools down sufficiently, the thermal switch will close, allowing power into the instrument.
Instruments without thermal cutoff switches are the exception. They are the 511, 512, 513, 514, 517 (serial numbers 1739 and below), and possibly others.
Certainly all of the 500-series scopes that take letter-series and 1-series plug-ins have thermal cutoff switches. This also includes the 127, 132, and 133 powered plug-in enclosures.
The 310A has an unusually high 165°F (74°C) thermal cutoff. This is due to the fact that the 310A is a compact tube scope without a fan.