067-0500-00: Difference between revisions
(New page: The Tektronix 067-0500-00 (also known as 3M1) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in 560-series scopes. It consists of a Schm...) |
No edit summary |
||
Line 10: | Line 10: | ||
After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted | After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted | ||
to produce clean transient response in the normalized scope. | to produce clean transient response in the normalized scope. | ||
The manual is here: [[http://bama.edebris.com/download/tek/067-0500-00/Tek%20067-0500-00.pdf]] |
Revision as of 07:34, 18 July 2010
The Tektronix 067-0500-00 (also known as 3M1) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in 560-series scopes. It consists of a Schmitt trigger that is driven from a single-ended input coming from a BNC connector on the front panel. The Schmitt trigger produces balanced output signals with controlled source impedance. The normalizing capacitor is adjusted using a procedure similar to adjusting probe compensation. When adjusted correctly, the voltage division ratio formed by the plug-in's output impedance and the mainframe's input impedance should be a real value which does not change across the specified frequency range of the scope. After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted to produce clean transient response in the normalized scope.
The manual is here: [[1]]