2215: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 12: Line 12:
* [[Media:tek_2215_service.pdf|Tektronix 2215 Manual (PDF, OCR)]]
* [[Media:tek_2215_service.pdf|Tektronix 2215 Manual (PDF, OCR)]]
* [[Media:Tek_2215A_Schematics_11x17.pdf|Tektronix 2215 Schematics (PDF, OCR)]]
* [[Media:Tek_2215A_Schematics_11x17.pdf|Tektronix 2215 Schematics (PDF, OCR)]]
* [[Media:070-4735-00.pdf|Tektronix 2215A Manual (PDF, OCR)]]
<small>
<small>
'''Modifications '''
'''Modifications '''

Revision as of 17:04, 18 July 2018

Manuals – Specifications – Links – Pictures


The Tektronix 2215 is a portable dual-trace 60 MHz scope from the 2000-series of low-cost analog scopes.

There is also a 2215A (same specifications as 2215 with improved CRT intensity and bandwidth limit ?)

Key Specifications

     — Vertical
Deflection 2 mV to 10 V per division, 1-2-5
Rise time < 5.8 ns (calculated)
Bandwidth DC to >60 MHz, 20 mV/Div to 10 V/Div, 0 °C to +40 °C; >50 MHz in 2 mV/Div to 10 mV/Div ranges or from +40 °C to +50 °C
Input impedance 1 MΩ ±2% // 30 ±3 pF
Maximum Safe Input Voltage 400 V (dc + peak AC) or 800 Vp-p AC to 1 kHz or less
CMRR At least 10:1 at 10 MHz
     — Horizontal
Timebase 50 ns/Div to 500 ms/Div, 1-2-5,  ×10 magnifier, i.e. down to 5 ns/Div
     — Trigger
Sensitivity 0.4 Div internal / 50 mV external up to 2 MHz; 1.5 Div internal / 250 mV external up to 60 MHz
AC coupled trigger ≤10 Hz (-3 dB)
External trigger input max. 400 V (DC+peak AC) or 800 Vp-p AC to 1kHz or less; impedance 1 MΩ ±2% // 30 ±3 pF
Trigger modes Auto, Norm, TV Field
     — Miscellaneous
Features
  • X-Y operation
  • Probe adjust (0.5V at 1kHz)
Z axis input (intensity); DC to 5 MHz
CRT display area : 80 × 100 mm²
Weight 6.1 kg; 7.6 kg with front panel cover, accessories and pouch
Dimensions With feet and handle: 137 x 361 x 511 mm³

Pictures