1M1: Difference between revisions

43 bytes added ,  17 March 2012
no edit summary
No edit summary
No edit summary
Line 1: Line 1:
The [[TU-7]]/1M1/[[067-0521-00]] is a letter series plug-in used for calibrating [[500-series scopes]].   
The [[TU-7]]/1M1/[[067-0521-00]] is a [[Letter-series_and_1-series_plug-ins|letter series plug-in]]
It is a multifunction replacement for the [[P|Type P Pulser]], the [[TU-2|TU-2 Test Load]], and the [[EP-53|EP-53 Gain Set Adaptor]].   
used for calibrating [[500-series scopes]].   
It is a multifunction replacement for the [[P|Type P Pulser]],  
the [[TU-2|TU-2 Test Load]], and the [[EP-53|EP-53 Gain Set Adaptor]].   
It is used for calibrating the mainframe's step response and gain,  
It is used for calibrating the mainframe's step response and gain,  
and checking dual-trace operation, common mode rejection, and power supply regulation.   
and checking dual-trace operation, common mode rejection,  
and power supply regulation.   
(500-series mainframes are designed to have 10 div/volt response to differential signals from the plug-in, and zero response to common-mode signals.)
(500-series mainframes are designed to have 10 div/volt response to differential signals from the plug-in, and zero response to common-mode signals.)