Revision as of 11:10, 12 August 2021
The Tektronix P6201 is an active 900 MHz ×1 FET probe.
It can be powered from 7000 series mainframes having LEMO S-series connectors for probe power, or compatible units such as the 1101.
The probe has a readout pin and the attenuation of the supplied attenuator plug-on tips is accurately encoded.
Key Specifications
Bandwidth | 900 MHz |
---|---|
Rise time | 0.39 ns max. |
Attenuation | × 1, optional × 10 and × 100 plug-on tips (010-0376-00, 010-0377-00) |
Dynamic range | ±0.6 V with ±5.6 V DC offset (× 1), ±6 V with ±56 V DC offset (× 10), ±60 V with ±200 V DC offset (× 100) |
Input impedance | 100 kΩ // 3 pF (× 1), 1 MΩ // 1.5 pF (× 10/100) |
Output impedance | 25 Ω (with internal termination selected) |
Intended for | 7000 series scopes, 485, 475 |
Pictures
-
P6201 kit
-
-
-
-
Circuitry in probe head
-
Lemo connector
-
-
-
-