Talk:AM503: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
Line 12: | Line 12: | ||
|- | |- | ||
! | ! | ||
! [[AM503]] !! [[AM503A]] !! [[AM503B]] !! [[AM5030]] !! [[11A16]] !! [[CT-4]] !! Max. current | ! [[AM503]] !! [[AM503A]] !! [[AM503B]] !! [[AM5030]] !! [[11A16]] !! [[CT-4]] !! Max. current !! Cable | ||
|- | |- | ||
! [[A6302]]/[[P6302]] | ! [[A6302]]/[[P6302]] | ||
Line 22: | Line 22: | ||
|align="center"| ✓ | |align="center"| ✓ | ||
| 20 A (DC + peak AC)<br />50 A peak, ≤100 A · μs | | 20 A (DC + peak AC)<br />50 A peak, ≤100 A · μs | ||
|align="center"| 2 m | |||
|- | |- | ||
! [[A6302XL]] | ! [[A6302XL]] | ||
Line 31: | Line 32: | ||
|align="center"| ✓ | |align="center"| ✓ | ||
| 20 A (DC + peak AC)<br />50 A peak, ≤100 A · μs | | 20 A (DC + peak AC)<br />50 A peak, ≤100 A · μs | ||
|align="center"| 8 m | |||
|- | |- | ||
! [[A6303]]/[[P6303]] | ! [[A6303]]/[[P6303]] | ||
Line 40: | Line 42: | ||
|align="center"| – | |align="center"| – | ||
| 100 A (DC + peak AC)<br/>500 A peak, ≤10 kA · μs | | 100 A (DC + peak AC)<br/>500 A peak, ≤10 kA · μs | ||
|align="center"| 2 m | |||
|- | |- | ||
! [[A6303XL]] | ! [[A6303XL]] | ||
Line 49: | Line 52: | ||
|align="center"| – | |align="center"| – | ||
| 100 A (DC + peak AC)<br/>500 A peak, ≤10 kA · μs | | 100 A (DC + peak AC)<br/>500 A peak, ≤10 kA · μs | ||
|align="center"| 8 m | |||
|- | |- | ||
! [[A6304XL]] | ! [[A6304XL]] | ||
Line 58: | Line 62: | ||
|align="center"| – | |align="center"| – | ||
| 500 A (DC + peak AC)<br/>700 A peak, ≤400 kA · μs | | 500 A (DC + peak AC)<br/>700 A peak, ≤400 kA · μs | ||
|align="center"| 8 m | |||
|- | |- | ||
! [[A6312]] | ! [[A6312]] | ||
Line 67: | Line 72: | ||
|align="center"| ✓ | |align="center"| ✓ | ||
| 20 A (DC + peak AC)<br/>50 A peak, ≤100 A · μs | | 20 A (DC + peak AC)<br/>50 A peak, ≤100 A · μs | ||
|align="center"| 2 m | |||
|- | |- | ||
|} | |} |
Revision as of 05:35, 8 January 2024
Are the AM503, AM503A, AM503B, AM5030 and 11A16 all electrically compatible with the same set of probes? Kurt (talk) 07:18, 5 January 2024 (PST)
- Good question ... what I can tell so far:
- The 11A16 manual lists only the A6302/P6302 and A6303/P6302 probes, as does the original AM503's. The A6312 came out later, its manual refers to all AM5x models (but not the 11A16).
- The A6304XL works only with the AM503B and AM5030, according to Media:070-8939-03.pdf.
- Media:070-3905-05.pdf says, The A6302XL can only be used with the AM503B and AM5030, and will not be recognized by other current probe amplifiers.
- Media:070-3906-06.pdf says, The A6303XL can only be used with the AM503B and AM5030, and will not be recognized by other current probe amplifiers.
- --Peter (talk) 17:43, 6 January 2024 (PST)
From the various manuals, I compiled the matrix below. --Peter (talk) 04:30, 8 January 2024 (PST)
AM503 | AM503A | AM503B | AM5030 | 11A16 | CT-4 | Max. current | Cable | |
---|---|---|---|---|---|---|---|---|
A6302/P6302 | 50 MHz | 50 MHz | 50 MHz | 50 MHz | 50 MHz 10.5 A |
✓ | 20 A (DC + peak AC) 50 A peak, ≤100 A · μs |
2 m |
A6302XL | – | – | 17 MHz | 17 MHz | – | ✓ | 20 A (DC + peak AC) 50 A peak, ≤100 A · μs |
8 m |
A6303/P6303 | 15 MHz | 15 MHz | 15 MHz | 15 MHz | 15 MHz | – | 100 A (DC + peak AC) 500 A peak, ≤10 kA · μs |
2 m |
A6303XL | – | – | 10 MHz | 10 MHz | – | – | 100 A (DC + peak AC) 500 A peak, ≤10 kA · μs |
8 m |
A6304XL | – | – | 2 MHz | 2 MHz | – | – | 500 A (DC + peak AC) 700 A peak, ≤400 kA · μs |
8 m |
A6312 | 100 MHz | 80 MHz | 100 MHz | 100 MHz | – | ✓ | 20 A (DC + peak AC) 50 A peak, ≤100 A · μs |
2 m |