The Tektronix 517 is a 50MHz scope [[introduced in 1950]] or [[introduced in 1951|1951]].  The power supply is an external box, like the [[507]], [[551]] and [[555]].   
The Tektronix 517 is a 50MHz scope [[introduced in 1950]] or [[introduced in 1951|1951]].  The power supply is an external box, like the [[507]], [[551]] and [[555]].   
The 517 does not take plug-ins.  There is also a 517A.   
The 517 does not take plug-ins.  There is also a 517A.
 
== Historic Context ==  
The 517 closely resembles a scope that Tektronix developed under a contract  
The 517 closely resembles a scope that Tektronix developed under a contract  
for the US military during 1949 and 1950.  (See link below.)
for the US military during 1949 and 1950.  (See link below.)
Line 9: Line 11:
</blockquote>
</blockquote>


== External Power Supply ==
The external power supply and oscilloscope are connected by a removable cable
The external power supply and oscilloscope are connected by a removable cable
with a Jones plug on one end and a Jones socket on the other end.   
with a Jones plug on one end and a Jones socket on the other end.   
Line 33: Line 36:
-150V on pin 8.   
-150V on pin 8.   


== Distributed Amplifiers ==
The 517 makes extensive use of the [[distributed amplifier]] concept.   
The 517 makes extensive use of the [[distributed amplifier]] concept.   
Originally the [[5XP]] CRT was used,  
Originally the [[5XP]] CRT was used,  
Line 67: Line 71:
* Gain Stage 4: single-ended, common-cathode, [[6AG7]] pentode
* Gain Stage 4: single-ended, common-cathode, [[6AG7]] pentode


== High Voltage Power Supply ==
The 517 uses 24kV total acceleration voltage on the CRT.  This is generated by the Type 420 High Voltage
The 517 uses 24kV total acceleration voltage on the CRT.  This is generated by the Type 420 High Voltage
Power Supply subsystem in the 517.  It uses a 1.8kHz oscillator to produce the high voltage, unlike
Power Supply subsystem in the 517.  It uses a 1.8kHz oscillator to produce the high voltage, unlike
other Tek scopes which use HV oscillators in the ultrasonic range.
other Tek scopes which use HV oscillators in the ultrasonic range.
The 517 has a switch on the front panel that selects between normal vertical sensitivity,
which uses the full 24kV acceleration voltage, and "X2" mode, which drops the acceleration
voltage to 12kV for a doubling of the vertical sensitivity.  The control works by
switching the the voltage division ratio of the feedback to the error amplifier in the HV supply.
   
   
The power consumption of a 517 is 1250 watts.  The indicator unit weighs 76 pounds and the external
The power consumption of a 517 is 1250 watts.  The indicator unit weighs 76 pounds and the external
power supply weighs 72 pounds.   
power supply weighs 72 pounds.   


== CRTs Used in the 517 ==
517 units with serial numbers 101 through 925 use the DuMont 5XP CRT.   
517 units with serial numbers 101 through 925 use the DuMont 5XP CRT.   
Later units use a Tek-made CRT, the T517PxH/T54PxH, where "x" designates
Later units use a Tek-made CRT, the T517PxH/T54PxH, where "x" designates
Line 79: Line 89:
6.6kV, 13.3kV, and 20kV.  The Tek-made CRT has just one anode connection, 20kV,
6.6kV, 13.3kV, and 20kV.  The Tek-made CRT has just one anode connection, 20kV,
and has 15 V/cm sensitivity at 24kV total acceleration voltage.
and has 15 V/cm sensitivity at 24kV total acceleration voltage.
The 517 has a switch on the front panel that selects between normal vertical sensitivity,
which uses the full 24kV acceleration voltage, and "X2" mode, which drops the acceleration
voltage to 12kV for a doubling of the vertical sensitivity.  The control works by
switching the the voltage division ratio of the feedback to the error amplifier in the HV supply.


The 517 might be the only Tektronix instrument to have a part made of wood.
The 517 might be the only Tektronix instrument to have a part made of wood.
Line 89: Line 94:
the probe power.
the probe power.


== Comparison of 517 with 547 ==
In 1965, following the introduction of the 50MHz [[547]], the
In 1965, following the introduction of the 50MHz [[547]], the
517 was discontinued.  The 547 uses a non-distributed solid-state vertical
517 was discontinued.  The 547 uses a non-distributed solid-state vertical