THS720: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
(pictures)
Line 1: Line 1:
[[Image:Ths720std-comp.jpg|thumb|400px|right|Tektronix THS720 Std probe compensation]]
[[Image:Ths720std-comp.jpg|thumb|400px|right|Tektronix THS720 Std probe compensation]]


The Tektronix THS720 is a portable 100 MHz 500 MS/s (each channel) two
The Tektronix THS720 is a portable 100 MHz bandwidth 500 MS/s (each channel) two
channel digital oscilloscope.  There were two variants, the THS720 STD
channel digital oscilloscope.  There were two variants, the THS720 STD
and the THS720A.  The primary difference is that the THS720A will trigger
and the THS720A.  The primary difference is that the THS720A will trigger
from the DMM input.
from the DMM input.
==Operation==
Due to the isolated inputs, the probe ground must be connected even when
compensating the probes with the "PROBE COMP" output.  It can be used with
any BNC probe but the scale factor pin found on many Tektronix probes is
not used.  The vertical menu has a scale factor entry (1X, 10X, 100X, 1000X)
which can also be used for direct BNC inputs.
Recommended probes are the [[P5102]] and the [[P6113B]], both are 10X probes
with protected grounds for isolated measurements.


==Internals==
==Internals==
Line 22: Line 33:
<gallery>
<gallery>
File:Ths720std-comp.jpg|THS720 Std displaying probe compensation waveform
File:Ths720std-comp.jpg|THS720 Std displaying probe compensation waveform
File:ths720std-backpcb.jpg|THS720 back of PCB
File:ths720std-toppcb.JPG|THS720 std top of main PCB
File:ths720std-mcudmm.JPG|THS720 std main board MCU and DMM
File:ths720std-powerbnc.JPG|THS720 std main board BNC inputs and power supply
File:ths720std-backpcb.jpg|THS720 std back of main PCB
</gallery>
</gallery>


[[Category:THS series scopes]]
[[Category:THS series scopes]]

Revision as of 18:15, 20 October 2014

Tektronix THS720 Std probe compensation

The Tektronix THS720 is a portable 100 MHz bandwidth 500 MS/s (each channel) two channel digital oscilloscope. There were two variants, the THS720 STD and the THS720A. The primary difference is that the THS720A will trigger from the DMM input.

Operation

Due to the isolated inputs, the probe ground must be connected even when compensating the probes with the "PROBE COMP" output. It can be used with any BNC probe but the scale factor pin found on many Tektronix probes is not used. The vertical menu has a scale factor entry (1X, 10X, 100X, 1000X) which can also be used for direct BNC inputs.

Recommended probes are the P5102 and the P6113B, both are 10X probes with protected grounds for isolated measurements.

Internals

Primary microprocessor is the Motorola/Freescale MC68331CFC16. There is also an Intel flash E28F008SA120 for setup, cal and waveform storage.

Repair Issues

The service manual shows how to remove the calibration protection jumper. The DMM calibration step requires a good calibrator capable of doing 850 VDC, and 600 VAC at 500 Hz.

See THS700.