067-0500-00: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 14: Line 14:
==Manuals==
==Manuals==
* [http://w140.com/tek_067-0500-00.pdf Tektronix 067-0500-00 Manual (PDF)]
* [http://w140.com/tek_067-0500-00.pdf Tektronix 067-0500-00 Manual (PDF)]
* [[Media:062-0806-00.pdf|Tektronix 067-0500-00 Manual (PDF, OCR)]]


==Pictures==
==Pictures==

Revision as of 11:06, 11 October 2018

The Tektronix 067-0500-00 (also known as 3M1) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in 560-series scopes.

It consists of a Schmitt trigger that is driven from a single-ended input coming from a BNC connector on the front panel. The Schmitt trigger produces balanced output signals with controlled source impedance. The normalizing capacitor is adjusted using a procedure similar to adjusting probe compensation. When adjusted correctly, the voltage division ratio formed by the plug-in's output impedance and the mainframe's input impedance should be a real value which does not change across the specified frequency range of the scope. After the scope is normalized using the 067-0500-00, other plug-ins can be adjusted to produce clean transient response in the normalized scope.

Manuals

Pictures