The S-3 probe contains a compensated attenuator before the sampling bridge. The [[P6038]] does not. | The S-3 probe contains a compensated attenuator before the sampling bridge. The [[P6038]] does not. | ||
The S-3 probe has a JFET buffer between the holding cap and the probe cable. The P6038 does not. | The S-3 probe has a JFET buffer between the holding cap and the probe cable. The P6038 does not. | ||
Both use the [[152-0368-00]] sampling bridge. | |||
== Repair of Sampling Diodes == | == Repair of Sampling Diodes == | ||
Individuals have reported success in replacing the S-3 sampling diodes with Agilent matched pair Schottky diodes. | Individuals have reported success in replacing the S-3 sampling diodes with Agilent matched pair Schottky diodes. |