DM501: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
Line 14: Line 14:
{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | DC V ranges | 200 mV, 2 V, 20 V, 200 V, 1000 V }}
{{Spec | DC V ranges | 200 mV, 2 V, 20 V, 200 V, 1000 V }}
{{Spec | DC V accuracy | 0.05% of reading + 0.01% FS in 2 V to 200 V ranges @ +18 °C to +28 °C }}
{{Spec | DC V accuracy | DM501: 0.1% of reading + 0,01% FS - DM501A: 0.05% of reading + 0.01% FS in 2 V to 200 V ranges @ +18 °C to +28 °C }}
{{Spec | DC V input resistance | 10 MΩ }}
{{Spec | DC V input resistance | 10 MΩ }}
{{Spec | AC V ranges | 200 mV, 2 V, 20 V, 200 V, 500 V }}
{{Spec | AC V ranges | 200 mV (DM501A only), 2 V, 20 V, 200 V, 500 V }}
{{Spec | AC V accuracy | 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C }}
{{Spec | AC V accuracy |DM501: 0.7% of reading + 0.01% FS - DM501A: 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C }}
{{Spec | AC V input resistance | 10 MΩ // 160 pF }}
{{Spec | AC V input resistance |DM501: 10 MΩ // 70pF - DM501A: 10 MΩ // 160 pF }}
{{Spec | dB ranges (DM501A only) | −40, −20, 0, +20, +40 dB }}
{{Spec | dB ranges (DM501A only) | −40, −20, 0, +20, +40 dB }}
{{Spec | Ω ranges | 200 Ω, 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20  MΩ }}
{{Spec | Ω ranges | 200 Ω (DM501A only), 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20  MΩ }}
{{Spec | Ω accuracy | 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C }}
{{Spec | Ω accuracy | DM501: 0.3% + 0.01% FS (2 kΩ to 2 MΩ) - DM501A: 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C }}
{{Spec | DC A ranges | 200 μA, 2 mA, 20 mA, 200 mA, 2 A }}
{{Spec | DC A ranges | 200 μA (DM501A only), 2 mA, 20 mA, 200 mA, 2 A }}
{{Spec | DC A accuracy | 0.2% of reading + 0.015% FS @ +18 °C to +28 °C }}
{{Spec | DC A accuracy |DM501: 0.2% of reading + 0.05% FS - DM501A: 0.2% of reading + 0.015% FS @ +18 °C to +28 °C }}
{{Spec | AC A ranges | 200 μA, 2 mA, 20 mA, 200 mA, 2 A }}
{{Spec | AC A ranges | 200 μA (DM501A only), 2 mA, 20 mA, 200 mA, 2 A }}
{{Spec | AC A accuracy | 0.6% of reading + 0.05% FS, 20 Hz to 10 kHz @ +18 °C to +28 °C }}
{{Spec | AC A accuracy |DM501: 0.6% + 0.01% FS - DM501A: 0.6% of reading + 0.05% FS, 40 Hz (20 Hz DM501A) to 10 kHz @ +18 °C to +28 °C }}
{{Spec | Temperature range | −62 °C to +240 °C }}
{{Spec | Temperature range | −62 °C to +240 °C }}
{{Spec | Temperature accuracy | ±2 °C from -62 °C to +150 °C with probe calibrated to instrument, @ +18 °C to +28 °C ambient }}
{{Spec | Temperature accuracy |DM501: ±1.5°C from -55°C to +125°C - DM501A: ±2 °C from -62 °C to +150 °C with probe calibrated to instrument, @ +18 °C to +28 °C ambient }}
{{EndSpecs}}
{{EndSpecs}}



Revision as of 04:33, 15 November 2020

Tektronix DM501
4½ digit True RMS multimeter
Tektronix DM501

Produced from 1972 to 1984

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix DM501 is a 4½ digit True RMS multimeter plug-in for the TM500 system.


It also has a temperature input, requiring the P6058 or P6430 temperature probe (DM501A: P6601 PT-100 probe).

The DM501 MOD718D variant replaced the temperature range with a 200 mV input range providing 10 μV resolution (see Physical Measurements Handbook).

Key Specifications

DC V ranges 200 mV, 2 V, 20 V, 200 V, 1000 V
DC V accuracy DM501: 0.1% of reading + 0,01% FS - DM501A: 0.05% of reading + 0.01% FS in 2 V to 200 V ranges @ +18 °C to +28 °C
DC V input resistance 10 MΩ
AC V ranges 200 mV (DM501A only), 2 V, 20 V, 200 V, 500 V
AC V accuracy DM501: 0.7% of reading + 0.01% FS - DM501A: 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C
AC V input resistance DM501: 10 MΩ // 70pF - DM501A: 10 MΩ // 160 pF
dB ranges (DM501A only) −40, −20, 0, +20, +40 dB
 Ω ranges 200 Ω (DM501A only), 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20  MΩ
 Ω accuracy DM501: 0.3% + 0.01% FS (2 kΩ to 2 MΩ) - DM501A: 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C
DC A ranges 200 μA (DM501A only), 2 mA, 20 mA, 200 mA, 2 A
DC A accuracy DM501: 0.2% of reading + 0.05% FS - DM501A: 0.2% of reading + 0.015% FS @ +18 °C to +28 °C
AC A ranges 200 μA (DM501A only), 2 mA, 20 mA, 200 mA, 2 A
AC A accuracy DM501: 0.6% + 0.01% FS - DM501A: 0.6% of reading + 0.05% FS, 40 Hz (20 Hz DM501A) to 10 kHz @ +18 °C to +28 °C
Temperature range −62 °C to +240 °C
Temperature accuracy DM501: ±1.5°C from -55°C to +125°C - DM501A: ±2 °C from -62 °C to +150 °C with probe calibrated to instrument, @ +18 °C to +28 °C ambient

Pictures

DM501