Tektronix 2247A
100 MHz portable 4-channel scope
Tektronix 2247A

Produced from 1989 to 1995

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 2247A is a 100 MHz four-channel analog oscilloscope with a built-in 200 MHz counter/timer. (It is unknown if a 2247 non-A model was ever built.) Channels 1-2 are fully-functional, whereas 3-4 are limited in both deflection factor and coupling, as they are intended mainly for digital and trigger signals.

The 2247A includes the same features of the 2246 plus a Counter/Timer (C/T) integrated into the trigger system to provide many automated counting and timing measurements of frequency, period, width, totalize, frequency ratio, gated frequency, gated period, gated width, gated events, delta time, one-over-delta time, phase, rise/fall time, and propagation delay, using an internal or external frequency standard.

Key Specifications

Bandwidth 100 MHz
Deflection CH 1, CH 2: 2 mV/div to 5 V/div in 1-2-5 seq; CH 3, CH 4: 0.1 V/div and 0.5 V/div, all ±2%; BW Limit: 20 MHz ±15%
Vertical Operating Modes CH1,2,3,4, CH2 Invert, Add, Alt, Chop (625 kHz)
Input Impedance 1 MΩ ±1.0% // 20 pF ±1 pF
Maximum Input Voltage 400 V (DC + peak AC) or 800 V (p-p ac at ≤ 10 kHz)
Sweep Speeds A Time Base: 0.5 s/div to 20 ns/div in 1-2-5 seq; B Time Base: 5 ms/div to 20 ns/div in 1-2-5 seq (×10 magnifier to 2 ns/div)
Trigger DC, Noise Reject, HF Reject, LF Reject, AC, TV Line, TV Field
Trigger Modes A Mode: Auto Level, Auto, Norm, TV Line, TV Field, Single Seq. B Mode: Runs After Delay, Auto Level, Norm, TV Line (From A Source)
Trigger Source A & B Vert, CH1, 2, 3, 4, Line
Voltmeter DC, +Peak, −Peak, PK-PK, Gated Volts, Smart Tracking Cursors: Volts, Gnd, Trigger Lvl
Counter/Timer Freq, Period, Width, Totalize, Freq Ratio, Sec, 1/Sec, Phase, Rise/Fall Time, Prop Delay Time
Cursors Time, 1/Time, Delta Time, 1/Delta Time, Delta Phase, Volts
Interface Ext Counter/Timer Base Input connector; Ext Z-Axis Input connector
Probes Two P6109 100 MHz ×10 probes, BNC connector with a readout pin
Power 100 W max, 90 − 250 VAC, 48 − 445 Hz