The Tektronix DM501 is a 4½ digit True RMS multimeter plug-in for the TM500 system.
It also has a temperature input, requiring the P6058 or P6430 temperature probe (DM501A: P6601 PT-100 probe).
The DM501 MOD718D variant replaced the temperature range with a 200 mV input range providing 10 μV resolution (see Physical Measurements Handbook).
Key Specifications
DC V ranges | 200 mV, 2 V, 20 V, 200 V, 1000 V |
---|---|
DC V accuracy | DM501: 0.1% of reading + 0.01% FS - DM501A: 0.05% of reading + 0.01% FS in 2 V to 200 V ranges @ +18 °C to +28 °C |
DC V input resistance | 10 MΩ |
AC V ranges | 200 mV (DM501A only), 2 V, 20 V, 200 V, 500 V |
AC V accuracy | DM501: 0.7% of reading + 0.01% FS - DM501A: 0.6% of reading + 0.05% FS in 0.2 V to 200 V ranges, 40 Hz to 10 kHz @ +18 °C to +28 °C |
AC V input resistance | DM501: 10 MΩ // 70pF - DM501A: 10 MΩ // 160 pF |
dB ranges (DM501A only) | −40, −20, 0, +20, +40 dB |
Ω ranges | 200 Ω (DM501A only), 2 kΩ, 20 kΩ, 200 kΩ, 2 MΩ, 20 MΩ |
Ω accuracy | DM501: 0.3% + 0.01% FS (2 kΩ to 2 MΩ) - DM501A: 0.15% of reading + 0.015% FS, 200 Ω to 2 MΩ ranges @ +18 °C to +28 °C |
DC A ranges | 200 μA (DM501A only), 2 mA, 20 mA, 200 mA, 2 A |
DC A accuracy | DM501: 0.2% of reading + 0.05% FS - DM501A: 0.2% of reading + 0.015% FS @ +18 °C to +28 °C |
AC A ranges | 200 μA (DM501A only), 2 mA, 20 mA, 200 mA, 2 A |
AC A accuracy | DM501: 0.6% + 0.01% FS - DM501A: 0.6% of reading + 0.05% FS, 40 Hz (20 Hz DM501A) to 10 kHz @ +18 °C to +28 °C |
Temperature range | −62 °C to +240 °C |
Temperature accuracy | DM501: ±1.5°C from -55°C to +125°C - DM501A: ±2 °C from -62 °C to +150 °C with probe calibrated to instrument, @ +18 °C to +28 °C ambient |