3M1
The Tektronix 3M1 (first designated the 067-050, then the 067-0050-00) is a calibration fixture for adjusting the normalizing capacitor behind the plug-in connector in 560-series scopes.
It consists of a Schmitt trigger that is driven from a single-ended input coming from a BNC connector on the front panel. The Schmitt trigger produces balanced output signals with controlled source impedance. The normalizing capacitor is adjusted using a procedure similar to adjusting probe compensation. When adjusted correctly, the voltage division ratio formed by the plug-in's output impedance and the mainframe's input impedance should be a real value which does not change across the specified frequency range of the scope. After the scope is normalized using the 3M1, other plug-ins can be adjusted to produce clean transient response in the normalized scope.
067-050
067-0050-00
3M1
-
3M1 Front
-
3M1 Circuit Board (note added resistor)
-
3M1 Schematic