Tektronix 1M1
Calibration plug-in
Tektronix 1M1

Compatible with 500-series scopes

Produced from 1964 to (?)

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix TU-7/1M1/067-0521-00 is a letter series plug-in used for calibrating 500-series scopes.

It is a multifunction replacement for the Type P Pulser, the TU-2 Test Load, and the EP-53 Gain Set Adaptor.

It is used for calibrating the mainframe's step response and gain, checking dual-trace operation, common mode rejection, and power supply regulation.

(500-series mainframes are designed to have 10 div/Volt response to differential signals from the plug-in, and zero response to common-mode signals.)

It contains a transistor multivibrator whose output is sharpened through Tektronix-made Gallium-Arsenide diodes (later production used silicon), a dual-trace circuit using 12AT7 and 6AL5 tubes, and various load resistors. It does not have the Type P's short operating life because it does not use a mercury switch.

This instrument went through several names. At its 1964 introduction, it was the TU-7. Some early Tektronix internal instruments for design, manufacturing, and calibration had TU (Test Unit) nomenclature. (Other examples include the TU-75B Variac, and the TU-50 sine/square/pulse generator.)

In 1965, it was renamed Type 1M1 to fit into the letter-series nomenclature. Later still, the 067- calibration fixture naming system took hold and it was rechristened 067-0521-00.

The instrument was also revised once - more pulse frequencies - and the name upgraded to 067-0521-01.

It does not appear - under any name - in any Tektronix catalog, but the 1964 catalog does list a cousin: Type 10/11M1, for the Type 647 scope.