Difference between revisions of "067-0502-00"

From TekWiki
Jump to navigation Jump to search
 
(13 intermediate revisions by 4 users not shown)
Line 1: Line 1:
The Tektronix 067-0502-00 is an instrument designed for calibrating oscilloscope calibrators.
+
[[File:067-0502-00 front.jpg | thumb | 350px |right]]
It is very similar to the [[067-0502-01]].  What are the differences?
+
The '''Tektronix 067-0502-00''' is an instrument designed for calibrating oscilloscope calibrators.
 +
It is very similar to the [[067-0502-01]].   
 +
 
 +
{{BeginSpecs}}
 +
{{Spec | Output Signal | DC+, DC-, Squarewave}}
 +
{{Spec | Amplitude Range | 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop}}
 +
{{Spec | Amplitude Accuracy | Within 0.25% into 1 MΩ load}}
 +
{{Spec | Squarewave Repetition Rate | 1 kHz within 20% }}
 +
{{Spec | Maximum Input Voltage | 100 V }}
 +
{{Spec | 100x Amplifier Gain Accuracy | ±1% }}
 +
{{Spec | 100x Amplifier Risetime | < 35  μs }}
 +
{{Spec | 100x Amplifier Output Range | ≥ ±20 V, ±5 mA }}
 +
{{Spec | 100x Amplifier Output Resistance | approx. 250 Ω}}
 +
{{EndSpecs}}
 +
 
 +
==Manuals==
 +
 
 +
* [[Media:062-0808-00.pdf|Tektronix 067-0502-00 Manual (PDF)]]
 +
* [[Media:Tek 067-0502-00.pdf|Tektronix 067-0502-00 Manual (PDF, OCR)]]
 +
 
 +
==Comparison with 067-0502-01==
  
 
Sherman Foy contributed the following information:
 
Sherman Foy contributed the following information:
 
<blockquote>
 
<blockquote>
TekWiki asks the question: what is the difference? I have both units and have done an extensive review.
+
TekWiki asks the question: What is the difference? I have both units and have done an extensive review.
  
Looking @ the front panels [TNX 4 pix] of the –00 & the –01, it seems that the  
+
Looking at the front panels [TNX 4 pix] of the –00 & the –01,  
–01 dropped the X100 Amplifier section.  The front panel is re-laid out, but what remains seems to be identical to the corresponding sections from the –00.  IOW, the –00 is a superset and the –01 is a subset of the -00.
+
it seems that the –01 dropped the × 100 Amplifier section.   
 +
The front panel is re-laid out,  
 +
but what remains seems to be identical to the corresponding sections from the –00.   
 +
In other words, the –00 provides a superset of the functionality of the -01.
  
Note that the –00 transformer [T310] has 6 windings and the –01 transformer [also T310] has 5 windings.  Some currents are different, as indicated.
+
Note that the –00 transformer [T310] has 6 windings  
 +
and the –01 transformer [also T310] has 5 windings.   
 +
Some currents are different, as indicated.
 +
 
 +
I have made my own notes for the project of working on the two.
 +
Using the schematics for organization, I arrived at the following table. 
 +
I am using a re-organization of the sheets, the block diagram of the –00 and this table:
  
I have made my own notes for the project of working on the two. Using the schematics for organization, I arrived @ the following table.  I am using a re-organization of the sheets, the block diagram of the –00 and this table:
 
</blockquote>
 
 
{| class="wikitable"
 
{| class="wikitable"
 
|-
 
|-
| Section
+
! scope="col"| Section
| -00
+
! scope="col"| -00
| -01
+
! scope="col"| -01
| Sheet in -00
+
! scope="col"| Sheet in -00
| Sheet in -01
+
! scope="col"| Sheet in -01
| Sheet in proposed consolidated drawing set
+
! scope="col"| Sheet in proposed consolidated drawing set
 
|-
 
|-
 
| Square Wave calibrator
 
| Square Wave calibrator
Line 29: Line 56:
 
| [s/b 1]
 
| [s/b 1]
 
|-
 
|-
| X100 Amplifier
+
| × 100 Amplifier
 
| Used in -00
 
| Used in -00
| Used in -01
+
| Deleted in -01
 
| 2
 
| 2
 
| --
 
| --
Line 51: Line 78:
 
|-
 
|-
 
| High voltage, B+
 
| High voltage, B+
| 285 VRMS, 35  mA 300 VDC, 225 VDC & 100 VDC
+
| 285 V<sub>RMS</sub>, 35 mA 300 V<sub>DC</sub>, 225 V<sub>DC</sub> &amp; 100 V<sub>DC</sub>
| 285 VRMS, 35  mA 300 VDC, 225 VDC & 100 VDC
+
| 125 V<sub>RMS</sub>, 2 mA 100 V<sub>DC</sub> only
 
| 3
 
| 3
 
| 2
 
| 2
 
| 3A/3B
 
| 3A/3B
 
|-
 
|-
| 6.3 VAC (Chopper)
+
| 6.3 V<sub>AC</sub> (Chopper)
 
| Used in -00
 
| Used in -00
| Used in -01, noted as 1.84A
+
| Used in -01, noted as 1.84 A
 
| 3
 
| 3
 
| 2
 
| 2
 
| 3A/3B
 
| 3A/3B
 
|-
 
|-
| 100 VRMS (-100 VDC)
+
| 100 V<sub>RMS</sub> (-100 V<sub>DC</sub>)
| Used in -00, 40 mA
+
| Used in -00, 40 mA
| Used in -01, 31 mA
+
| Used in -01, 31 mA
 
| 3
 
| 3
 
| 2
 
| 2
 
| 3A/3B
 
| 3A/3B
 
|-
 
|-
| 6.3 VAC (MV 6DJ8)
+
| 6.3 V<sub>AC</sub> (MV 6DJ8)
 
| Used in -00
 
| Used in -00
 
| Used in -01
 
| Used in -01
Line 78: Line 105:
 
| 3A/3B
 
| 3A/3B
 
|-
 
|-
| 330 VRMS (Floating Supply)
+
| 330 V<sub>RMS</sub> (Floating Supply)
 
| Used in -00
 
| Used in -00
 
| Used in -01
 
| Used in -01
Line 85: Line 112:
 
| 3A/3B
 
| 3A/3B
 
|-
 
|-
| 20 VRMS (X100 Filaments)
+
| 20 V<sub>RMS</sub> (× 100 Filaments)
 
| Used in -00
 
| Used in -00
 
| Not used in –01 T310
 
| Not used in –01 T310
Line 99: Line 126:
 
| [s/b 4]
 
| [s/b 4]
 
|}
 
|}
 +
</blockquote>
  
* [http://w140.com/Tek_067-0502-00.pdf Tektronix 067-0502-00 Manual (PDF)]
+
==Pictures==
 
 
 
<gallery>
 
<gallery>
 
File:067-0502-00 front.jpg
 
File:067-0502-00 front.jpg
Line 108: Line 135:
  
 
[[Category:Calibration fixtures]]
 
[[Category:Calibration fixtures]]
 +
[[Category:Pulse generators]]

Latest revision as of 22:52, 9 September 2020

067-0502-00 front.jpg

The Tektronix 067-0502-00 is an instrument designed for calibrating oscilloscope calibrators. It is very similar to the 067-0502-01.

Specifications

Output Signal DC+, DC-, Squarewave
Amplitude Range 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop
Amplitude Accuracy Within 0.25% into 1 MΩ load
Squarewave Repetition Rate 1 kHz within 20%
Maximum Input Voltage 100 V
100x Amplifier Gain Accuracy ±1%
100x Amplifier Risetime < 35  μs
100x Amplifier Output Range ≥ ±20 V, ±5 mA
100x Amplifier Output Resistance approx. 250 Ω

Manuals

Comparison with 067-0502-01

Sherman Foy contributed the following information:

TekWiki asks the question: What is the difference? I have both units and have done an extensive review.

Looking at the front panels [TNX 4 pix] of the –00 & the –01, it seems that the –01 dropped the × 100 Amplifier section. The front panel is re-laid out, but what remains seems to be identical to the corresponding sections from the –00. In other words, the –00 provides a superset of the functionality of the -01.

Note that the –00 transformer [T310] has 6 windings and the –01 transformer [also T310] has 5 windings. Some currents are different, as indicated.

I have made my own notes for the project of working on the two. Using the schematics for organization, I arrived at the following table. I am using a re-organization of the sheets, the block diagram of the –00 and this table:

Section -00 -01 Sheet in -00 Sheet in -01 Sheet in proposed consolidated drawing set
Square Wave calibrator Used in both Used in both 1 1 [s/b 1]
× 100 Amplifier Used in -00 Deleted in -01 2 -- [s/b 2]
Power Supply -00 has more sections Than –01 3 2 [s/b 3A for -00 & 3B for -01]
T310 P/N 120-0442-00 P/N 120-0615-00 3 2 3A/3B
High voltage, B+ 285 VRMS, 35 mA 300 VDC, 225 VDC & 100 VDC 125 VRMS, 2 mA 100 VDC only 3 2 3A/3B
6.3 VAC (Chopper) Used in -00 Used in -01, noted as 1.84 A 3 2 3A/3B
100 VRMS (-100 VDC) Used in -00, 40 mA Used in -01, 31 mA 3 2 3A/3B
6.3 VAC (MV 6DJ8) Used in -00 Used in -01 3 2 3A/3B
330 VRMS (Floating Supply) Used in -00 Used in -01 3 2 3A/3B
20 VRMS (× 100 Filaments) Used in -00 Not used in –01 T310 3 2 3A/3B
Floating power supply Used in both; drives Square Wave MV Used in both; drives Square Wave MV 4 3 [s/b 4]

Pictures