The Tektronix 067-0502-00 is an instrument designed for calibrating oscilloscope calibrators. | [[File:067-0502-00 front.jpg | thumb | 350px |right]] | ||
It is very similar to the [[067-0502-01]]. | The '''Tektronix 067-0502-00''' is an instrument designed for calibrating oscilloscope calibrators. | ||
It is very similar to the [[067-0502-01]]. | |||
{{BeginSpecs}} | |||
{{Spec | Output Signal | DC+, DC-, Squarewave}} | |||
{{Spec | Amplitude Range | 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop}} | |||
{{Spec | Amplitude Accuracy | Within 0.25% into 1 MΩ load}} | |||
{{Spec | Squarewave Repetition Rate | 1 kHz within 20% }} | |||
{{Spec | Maximum Input Voltage | 100 V }} | |||
{{Spec | 100x Amplifier Gain Accuracy | ±1% }} | |||
{{Spec | 100x Amplifier Risetime | < 35 μs }} | |||
{{Spec | 100x Amplifier Output Range | ≥ ±20 V, ±5 mA }} | |||
{{Spec | 100x Amplifier Output Resistance | approx. 250 Ω}} | |||
{{EndSpecs}} | |||
==Manuals== | |||
* [[Media:062-0808-00.pdf|Tektronix 067-0502-00 Manual (PDF)]] | |||
* [[Media:Tek 067-0502-00.pdf|Tektronix 067-0502-00 Manual (PDF, OCR)]] | |||
==Comparison with 067-0502-01== | |||
Sherman Foy contributed the following information: | Sherman Foy contributed the following information: | ||
Line 7: | Line 27: | ||
Looking at the front panels [TNX 4 pix] of the –00 & the –01, | Looking at the front panels [TNX 4 pix] of the –00 & the –01, | ||
it seems that the –01 dropped the | it seems that the –01 dropped the × 100 Amplifier section. | ||
The front panel is re-laid out, | The front panel is re-laid out, | ||
but what remains seems to be identical to the corresponding sections from the –00. | but what remains seems to be identical to the corresponding sections from the –00. | ||
Line 36: | Line 56: | ||
| [s/b 1] | | [s/b 1] | ||
|- | |- | ||
| | | × 100 Amplifier | ||
| Used in -00 | | Used in -00 | ||
| Deleted in -01 | | Deleted in -01 | ||
Line 58: | Line 78: | ||
|- | |- | ||
| High voltage, B+ | | High voltage, B+ | ||
| | | 285 V<sub>RMS</sub>, 35 mA 300 V<sub>DC</sub>, 225 V<sub>DC</sub> & 100 V<sub>DC</sub> | ||
| | | 125 V<sub>RMS</sub>, 2 mA 100 V<sub>DC</sub> only | ||
| 3 | | 3 | ||
| 2 | | 2 | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| 6. | | 6.3 V<sub>AC</sub> (Chopper) | ||
| Used in -00 | | Used in -00 | ||
| Used in -01, noted as 1. | | Used in -01, noted as 1.84 A | ||
| 3 | | 3 | ||
| 2 | | 2 | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| | | 100 V<sub>RMS</sub> (-100 V<sub>DC</sub>) | ||
| Used in -00, | | Used in -00, 40 mA | ||
| Used in -01, | | Used in -01, 31 mA | ||
| 3 | | 3 | ||
| 2 | | 2 | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| 6. | | 6.3 V<sub>AC</sub> (MV 6DJ8) | ||
| Used in -00 | | Used in -00 | ||
| Used in -01 | | Used in -01 | ||
Line 85: | Line 105: | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| | | 330 V<sub>RMS</sub> (Floating Supply) | ||
| Used in -00 | | Used in -00 | ||
| Used in -01 | | Used in -01 | ||
Line 92: | Line 112: | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| | | 20 V<sub>RMS</sub> (× 100 Filaments) | ||
| Used in -00 | | Used in -00 | ||
| Not used in –01 T310 | | Not used in –01 T310 | ||
Line 107: | Line 127: | ||
|} | |} | ||
</blockquote> | </blockquote> | ||
==Pictures== | |||
<gallery> | <gallery> | ||
File:067-0502-00 front.jpg | File:067-0502-00 front.jpg | ||
Line 115: | Line 135: | ||
[[Category:Calibration fixtures]] | [[Category:Calibration fixtures]] | ||
[[Category:Pulse generators]] |
Revision as of 22:52, 9 September 2020
The Tektronix 067-0502-00 is an instrument designed for calibrating oscilloscope calibrators. It is very similar to the 067-0502-01.
Key Specifications
Output Signal | DC+, DC-, Squarewave |
---|---|
Amplitude Range | 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop |
Amplitude Accuracy | Within 0.25% into 1 MΩ load |
Squarewave Repetition Rate | 1 kHz within 20% |
Maximum Input Voltage | 100 V |
100x Amplifier Gain Accuracy | ±1% |
100x Amplifier Risetime | < 35 μs |
100x Amplifier Output Range | ≥ ±20 V, ±5 mA |
100x Amplifier Output Resistance | approx. 250 Ω |
Manuals
Comparison with 067-0502-01
Sherman Foy contributed the following information:
TekWiki asks the question: What is the difference? I have both units and have done an extensive review.
Looking at the front panels [TNX 4 pix] of the –00 & the –01, it seems that the –01 dropped the × 100 Amplifier section. The front panel is re-laid out, but what remains seems to be identical to the corresponding sections from the –00. In other words, the –00 provides a superset of the functionality of the -01.
Note that the –00 transformer [T310] has 6 windings and the –01 transformer [also T310] has 5 windings. Some currents are different, as indicated.
I have made my own notes for the project of working on the two. Using the schematics for organization, I arrived at the following table. I am using a re-organization of the sheets, the block diagram of the –00 and this table:
Section -00 -01 Sheet in -00 Sheet in -01 Sheet in proposed consolidated drawing set Square Wave calibrator Used in both Used in both 1 1 [s/b 1] × 100 Amplifier Used in -00 Deleted in -01 2 -- [s/b 2] Power Supply -00 has more sections Than –01 3 2 [s/b 3A for -00 & 3B for -01] T310 P/N 120-0442-00 P/N 120-0615-00 3 2 3A/3B High voltage, B+ 285 VRMS, 35 mA 300 VDC, 225 VDC & 100 VDC 125 VRMS, 2 mA 100 VDC only 3 2 3A/3B 6.3 VAC (Chopper) Used in -00 Used in -01, noted as 1.84 A 3 2 3A/3B 100 VRMS (-100 VDC) Used in -00, 40 mA Used in -01, 31 mA 3 2 3A/3B 6.3 VAC (MV 6DJ8) Used in -00 Used in -01 3 2 3A/3B 330 VRMS (Floating Supply) Used in -00 Used in -01 3 2 3A/3B 20 VRMS (× 100 Filaments) Used in -00 Not used in –01 T310 3 2 3A/3B Floating power supply Used in both; drives Square Wave MV Used in both; drives Square Wave MV 4 3 [s/b 4]