Tektronix 067-0625-00 Peak to Peak Detector
{{Instrument Sidebar
|manufacturer=Tektronix
|class=Accessory
|model=067-0625-00
|summary=Peak-to-Peak Detector
|image=Tek 0625 1.JPG
|caption=Tektronix 067-0625-00
|introduced=(?)
|discontinued=(?)
|manuals=
* [[Media:062-1306-00.pdf|Tektronix 067-0625-00 Manual -00]] (PDF)
* [[Media:062-1306-01.pdf|Tektronix 067-0625-00 Manual -01]] (PDF)
* [[Media:Tek 067-0625-00 service cal procedure.pdf|Tektronix 067-0625 Service Calibration Procedure (OCR)]]
}}
The '''Tektronix 067-0625-00''' is a peak-to-peak detector designed to measure the output voltage flatness
of constant amplitude signal generators in the 50 Hz to 500 MHz range.
The input amplitude should be 1.2 V<sub>p-p</sub>.


* [http://w140.com/tek_067-0625-00.pdf Tektronix 067-0625-00 Manual (PDF)]
The input is a [[GR-874 connector]] terminated in 50 Ω.


There are two outputs, one for the positive peak voltage and one for the negative peak voltage.
The outputs are [[BNC]] females, intended for connection to high-impedance low-frequency
instrumentation.
Up to serial number B003240, the 067-0625-00 used 1.35 V mercury batteries for biasing the detector diodes.
Starting at serial number B003241, 1.5 V alkaline batteries were used.
{{BeginSpecs}}
{{Spec | Sensitivity | 10 ±1 mV/% input change at 1.2 V<sub>p-p</sub> }}
{{Spec | Temperature Range | 23&deg;C ±10&deg;C }}
{{Spec | Output | Into floating 2 MΩ load or differential oscilloscope with 1 MΩ inputs }}
{{EndSpecs}}
==Links==
* [https://www.youtube.com/watch?v=VVPHifwseJc Building a 067-0625-00 clone] by NFM @ YouTube
==Pictures==
<gallery>
<gallery>
File:Tek 0625 1.JPG
Tek 0625 1.JPG
File:Tek 0625 2.JPG
Tek 0625 2.JPG
File:Tek 067-0625-00 schem.png
Tek 067-0625-00 schem.png
</gallery>
</gallery>
[[Category:Test adapters]]