==DEFLECTION FACTORS AND ASSOCIATED ADDRESS TABLE==
==DEFLECTION FACTORS AND ASSOCIATED ADDRESS TABLE==
Deflection Factor
{| class=wikitable
50mV/div 20mV/div 10mV/div
|Deflection Factor
5mV/div 2mV/div 1mV/div
|CH1 Cal Address
|CH2 Cal Address
|-
|50mV/div  
|58
|186
|-
|20mV/div  
|57
|185
|-
|10mV/div
|56
|184
|-
|5mV/div  
|55
|183
|-
|2mV/div
|54
|182
|-
|1mV/div
|53
|181
|}
 
That should do it.
That should do it.
CH1 Cal Address
 
    58
Return the plug-in J611 jumper to the vertical position.
    57
Return the mainframe Cal Lock jumper to the disconnected state.
    56
 
    55
    54
    53
CH2 Cal Address
    186
    185
    184
    183
    182
    181
  Return the plugin disconnected state.
J611 jumper to the
vertical position. Return the mainframe Cal Lock jumper to the
With 12 new HFAdj cal constants and a new serial number, the instrument will probably come up without errors.  
With 12 new HFAdj cal constants and a new serial number, the instrument will probably come up without errors.  
If there are probe related errors,  
If there are probe related errors,  
Line 291: Line 305:
on the front panel with the instrument running and then removing it.  
on the front panel with the instrument running and then removing it.  
Simply recycling the power or shorting the probe ring may also do this.
Simply recycling the power or shorting the probe ring may also do this.
You can check that the serial number is correct by pressing the mainframe front panel “UTILITY” button and touching the “Ident” part of the screen.  
 
You can check that the serial number is correct by pressing  
the mainframe front panel “UTILITY” button and touching the “Ident” part of the screen.  
All the installed plugin serial numbers and firmware versions should be shown.
All the installed plugin serial numbers and firmware versions should be shown.