The '''Tektronix Type 130''' is a self-contained instrument that measures inductance and capacitance.   
|manufacturer=Tektronix
When measuring inductance, the 130 makes an LC oscillator using the device (inductance) under test
|model=130
and a capacitor internal to the 130.  When measuring capacitance, the 130 makes an LC oscillator using the device
|class=Multimeter
(capacitance) under test and an inductor internal to the 130.  The frequency of the resulting LC oscillator is
|series=
measured by mixing it down using a fixed reference frequency, generating a pulse at each zero-crossing, and
|summary=L/C Meter
integrating (low-pass filtering) the pulse train.  This produces a voltage that is proportional to the
|image=Tek_130-Models.jpeg
frequency difference between the LC oscillator and the reference oscillator.  This voltage is displayed
|caption=LC 130, SN 101-5000 vs. 5001-Up
on a d'Arsonval (moving coil) meter on the front panel of the 130.
|introduced=1959
|discontinued=1968
|designers=Cliff Moulton
|manuals=
* [[Media:070-231.pdf | Tektronix 130 Manual 070-231]]
* [https://vintagetek.org/wp-content/uploads/2020/09/Alan-Hampel-Tektronix-130-Instruction-Manual.pdf 070-231 Manual Restored by Alan Hampel @ vintagetek.org] (includes redrawn schematics)
* [[Media:070-0231-01.pdf | Tektronix 130 Manual 070-231-01]] (covers both cabinet types)
* [[Media:tek_type_130_factory_cal_proc.pdf|Tektronix 130 Factory Calibration Procedure]]
* [[Media:Tek 130 calibration outline.pdf|Tektronix 130 Calibration Outline]] (OCR)
}}
The '''Tektronix Type 130''' is a self-contained instrument, designed by [[Cliff Moulton]] and [[introduced in 1959]], that measures inductance and capacitance.   


The Type 130 was designed by [[Cliff Moulton]].
There are at least two versions of Type 130.
Early units (SN 101-5000) have a narrow boxy case that slides off the chassis, backlit meter, and are marked "TYPE 130 L,C METER".
At some point between SN 101 and SN 842, the case color changed from brown to blue.
Later units (SN 5001-Up) have a wider, rounded case with side panels that pop off, blue paint, non-backlit meter, and are marked "TYPE 130 L-C METER".


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Ranges | 3, 10, 30, 100 or 300 pF / 3, 10, 30, 100 or 300 μH full scale}}
{{Spec | Ranges | 3, 10, 30, 100 or 300 pF / 3, 10, 30, 100 or 300 μH full scale}}
{{Spec | Accuracy | 3% FS, calibratable to 1% FS }}
{{Spec | Accuracy | 3% FS, 1% FS with "careful" calibration using [[S-30|S-30 Delta Standard]]}}
{{Spec | Measurement voltage | <1 V for capacitors, <0.25 V for inductors, 120-140 kHz }}
{{Spec | Measurement voltage | <1 V for capacitors, <0.25 V for inductors, 120-140 kHz }}
{{Spec | Guard output | 250 Ω source impedance, can drive 200 pF }}
{{Spec | Guard output | 250 Ω source impedance, can drive 200 pF }}
Line 19: Line 32:
{{EndSpecs}}
{{EndSpecs}}


==Manuals==
==Links==
* [http://bama.edebris.com/download/tek/130/Tek-130_LC_Meter.pdf Tektronix 130 Manual (PDF)]
* Q+A: Type 130 L-C Meter and S-30 Delta Standards. [[Media:Service Scope 18 Feb 1963.pdf | Service Scope No. 18, Feb 1963]].
* [http://w140.com/tek_fcp/tek_type_130_factory_cal_proc.pdf Tektronix 130 Factory Calibration Procedure (PDF)]
* Silicon Chip, June 2020 – Vintage Workbench – Tektronix T-130 LC Meter:  [https://vintagetek.org/wp-content/uploads/2020/09/Silicon-Chip-June-2020-Vintage-Workbench-Tektronix-T-130-LC-Meter-part-one.pdf Part 1] / [https://vintagetek.org/wp-content/uploads/2020/09/Silicon-Chip-July-2020-Vintage-Workbench-Tektronix-T-130-LC-Meter-part-two.pdf Part 2] / [https://vintagetek.org/wp-content/uploads/2020/09/Silicon-Chip-August-2020-Vintage-Workbench-Tektronix-T-130-LC-Meter-part-three.pdf Part 3] (PDFs @ vintagetek.org)


==Internals==
The 130 contains an LC oscillator with a tank circuit formed by the device under test and a capacitor or inductor internal to the 130.
The oscillator's frequency is measured by mixing it down using a fixed reference frequency, generating a pulse at each zero-crossing,
and integrating (low-pass filtering) the pulse train.
This produces a voltage that is proportional to the frequency difference between the LC oscillator and the reference oscillator. 
This voltage is displayed on a d'Arsonval (moving coil) meter on the front panel of the 130.
==Pictures==
=== 130 External ===
<gallery>
Tek_130-Models.jpeg | 130 Variants
Tek_130-Early-front.jpeg | Tek 130 SN < 5001 Front
Tek_Type130_Front.jpg | Tek 130 130 SN > 5000 Front
Tek 130 early.jpg|Early (Brown Era) Type 130
Tek 130 early2.jpg|Early (Brown Era) Type 130
Martin 130 01.jpg
Martin 130 02.jpg
Martin 130 03.jpg
Tek_Type130_Back.jpg | Type 130 Rear
Martin 130 04.jpg|[[013-0001-00]] test adapter
</gallery>
=== 130 Internal ===
<gallery>
<gallery>
File:Tek 130 early.jpg|Early (Brown Era) Type 130
Tek_Type130_Inside-R.jpg | Type 130 Internal RHS
File:Tek 130 early2.jpg|Early (Brown Era) Type 130
Tek_130-Early-RHS.jpeg | Tek 130 SN < 5001 RHS
File:Martin 130 01.jpg
Tek_Type130_Inside-L.jpg | Type 130 Internal LHS
File:Martin 130 02.jpg
Tek_130-Early-LHS.jpeg | Tek 130 SN < 5001 LHS (w line plug mod)
File:Martin 130 03.jpg
Tek_Type130_Top.jpg  | Type 130 Top w/o Cover
File:Martin 130 04.jpg|[[013-001-00]]
Tek_Type130_Bottom.jpg | Type 130 Bottom
File:Martin 130 05.jpg
Tek_Type130_Inside-1.jpg |  Type 130 Internal #1
File:Martin 130 06.jpg
Tek_Type130_Inside-2.jpg |  Type 130 Internal #1
File:Martin 130 07.jpg
Tek_Type130_Live.jpg | Type 130 Internal Live
File:Martin 130 08.jpg
Martin 130 05.jpg
File:Martin 130 09.jpg
Martin 130 06.jpg
File:Martin 130 10.jpg
Martin 130 07.jpg
File:Martin 130 11.jpg
Martin 130 08.jpg
Martin 130 09.jpg
Martin 130 10.jpg
Martin 130 11.jpg
</gallery>
</gallery>
=== 130 SN 842 ===
At serial number 842, the shape is still narrow and boxy, but with a blue case.
<gallery>
Tek 130 sn842 1.jpg
Tek 130 sn842 2.jpg
Tek 130 sn842 3.jpg
Tek 130 sn842 4.jpg
Tek 130 sn842 5.jpg
Tek 130 sn842 6.jpg
Tek 130 sn842 7.jpg
</gallery>
{{Parts|130}}


[[Category:LCR meters]]
[[Category:LCR meters]]

Revision as of 05:32, 26 January 2023

Tektronix 130
L/C Meter
LC 130, SN 101-5000 vs. 5001-Up

Produced from 1959 to 1968

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix Type 130 is a self-contained instrument, designed by Cliff Moulton and introduced in 1959, that measures inductance and capacitance.

There are at least two versions of Type 130. Early units (SN 101-5000) have a narrow boxy case that slides off the chassis, backlit meter, and are marked "TYPE 130 L,C METER". At some point between SN 101 and SN 842, the case color changed from brown to blue. Later units (SN 5001-Up) have a wider, rounded case with side panels that pop off, blue paint, non-backlit meter, and are marked "TYPE 130 L-C METER".

Key Specifications

Ranges 3, 10, 30, 100 or 300 pF / 3, 10, 30, 100 or 300 μH full scale
Accuracy 3% FS, 1% FS with "careful" calibration using S-30 Delta Standard
Measurement voltage <1 V for capacitors, <0.25 V for inductors, 120-140 kHz
Guard output 250 Ω source impedance, can drive 200 pF
DUT connection UHF connector (DUT) + 4 mm jack (guard)

Links

Internals

The 130 contains an LC oscillator with a tank circuit formed by the device under test and a capacitor or inductor internal to the 130. The oscillator's frequency is measured by mixing it down using a fixed reference frequency, generating a pulse at each zero-crossing, and integrating (low-pass filtering) the pulse train. This produces a voltage that is proportional to the frequency difference between the LC oscillator and the reference oscillator. This voltage is displayed on a d'Arsonval (moving coil) meter on the front panel of the 130.

Pictures

130 External

130 Internal

130 SN 842

At serial number 842, the shape is still narrow and boxy, but with a blue case.


Some Parts Used in the 130

Part Part Number(s) Class Description Used in
0A2 154-0001-00 Gas Discharge Tube (Voltage regulator) 150 V voltage reference 130
6BE6 154-0025-00 Vacuum Tube (Pentode) heptode 130
6BH6 154-0026-00 154-0285-00 Vacuum Tube (Pentode) pentode 130 162 310 310A 512 517 517A 524 Nelson-Ross PSA-011
6BW7 154-0028-00 Vacuum Tube (Pentode) pentode 130
6U8 154-0033-00 Vacuum Tube (Triode/Pentode) triode-pentode combo 130 160 163 310 315 316 524 535 545 T
6X4 154-035 154-0035-00 Vacuum Tube (Dual Rectifier) dual rectifier 128 130 181 511A 512 517 517A 570