No edit summary |
||
(19 intermediate revisions by 5 users not shown) | |||
Line 5: | Line 5: | ||
caption=1S2 front view| | caption=1S2 front view| | ||
introduced=1967 | | introduced=1967 | | ||
discontinued= | discontinued=1973 | | ||
series=[[500-series scopes]]| | series=[[500-series scopes]]| | ||
manuals= | manuals= | ||
* [ | * [[Media:070-0889-00.pdf|1S2 manual (PDF, sn 1990 and up, OCR)]] | ||
* [http://www.pa4tim.nl/wp-content/uploads/2012/01/1S2calibrationPA4TIM.pdf 1S2 Calibration Guide by Fred Schneider, PA4TIM] | * [http://www.pa4tim.nl/wp-content/uploads/2012/01/1S2calibrationPA4TIM.pdf 1S2 Calibration Guide by Fred Schneider, PA4TIM] | ||
* [http://www.pa4tim.nl/?p=3016 1S2 Page on the Website of Fred Schneider, PA4TIM] | * [http://www.pa4tim.nl/?p=3016 1S2 Page on the Website of Fred Schneider, PA4TIM] | ||
* [[Media:Tek 1s2 fcp june 1966.pdf|Tektronix 1S2 Factory Calibration Procedure, June 1966]] | |||
* [[Media:Tek 1s2 fcp aug 1968.pdf|Tektronix 1S2 Factory Calibration Procedure, August 1968]] | |||
<!-- * [[Media:1S2_calibration_guide.pdf|1S2 Calibration Guide by Fred Schneider, PA4TIM (PDF, OCR)]] --> | |||
}} | }} | ||
The '''Tektronix 1S2''' is a [[sampling oscilloscope|sampling]]/TDR plug-in for [[500-series scopes]] | The '''Tektronix 1S2''' is a [[sampling oscilloscope|sampling]]/TDR plug-in for [[500-series scopes]], [[introduced in 1967]] and sold through 1972. | ||
Unlike the 1S1, | Like the [[1S1]], it is a complete sampling subsystem and uses the mainframe merely as a power supply and low-speed X-Y display. | ||
The 1S2 outputs the Y-axis signal on plug-in interface connector pins 1 and 3 like other 500-series plug-ins, and the X-axis signal on the front panel, via a banana connector. The X-axis sensitivity of the oscilloscope mainframe is expected to be 1 V/div. | |||
Unlike the 1S1, the 1S2 contains pulsers that allows it to perform time-domain reflectometry (TDR) measurements. | |||
The 1S2 has two 50 Ω pulsers, one of which is active at a time. One puts out a 250 mV pulse with 50 ps rise time. | The 1S2 has two 50 Ω pulsers, one of which is active at a time. One puts out a 250 mV pulse with 50 ps rise time. | ||
The other puts out a slower | The other puts out a slower 1 V pulse. The sampling bridge is a feed-through type. | ||
For TDR measurements, the signal from one of the internal pulsers is fed through the sampler, and out to | For TDR measurements, the signal from one of the internal pulsers is fed through the sampler, and out to the transmission line being tested. | ||
Returning pulses are detected as they pass back through the sampler, and they are reabsorbed by the 50 Ω impedance of the pulser. | Returning pulses are detected as they pass back through the sampler, and they are reabsorbed by the 50 Ω impedance of the pulser. | ||
Waveform output can be to the CRT of the mainframe and/or to a plotter or low-speed data acquisition device via banana connectors on the front of the 1S2. | Waveform output can be to the CRT of the mainframe and/or to a plotter or low-speed data acquisition device via banana connectors on the front of the 1S2. | ||
The 1S2 | The interface between the GR-874 connector and the printed circuit board is described in US Patent 3,426,311. | ||
{{MissingSpecs}} | |||
==Internals== | |||
The 1S2 contains an internal power supply producing regulated +19 V, -19 V, and -136 V outputs, | |||
derived from a power transformer whose primary voltage is the 6.3 V<sub>AC</sub> | |||
provided by the 500-series mainframe on pins 13 and 14 of the plug-in interface connector. | |||
The regulators use bipolar transistors for the feedback amplifiers and the output emitter-followers, and zener diodes as references. | |||
==Links== | ==Links== | ||
* [ | * [[Media:062-1244-00.pdf | Tektronix 062-1244-00: Time-Domain Reflectometry Measurement Concepts]], James A. Strickland, 1969. pages. PDF, OCR, 6.2 MB | ||
* [[Media:Service Scope 45 Aug 1967.pdf | Service Scope No. 45, August 1967]] | * [[Media:Service Scope 45 Aug 1967.pdf | Service Scope No. 45, August 1967 - Time-Domain Reflectometry Theory]] | ||
* [[Media:Service Scope 38 Jun 1966.pdf | Service Scope No. 38, June 1966 - Tektronix Type 1S2 Makes Reflectometry Easy]] | |||
* [http://w140.com/kurt/50_years_sampling.pdf 50 Years of RF and Microwave Sampling, Mark Kahrs] | * [http://w140.com/kurt/50_years_sampling.pdf 50 Years of RF and Microwave Sampling, Mark Kahrs] | ||
* [http://www.picosecond.com/objects/AN-04.pdf Picosecond Pulse Labs Application Note 4: TDR] | * [http://www.picosecond.com/objects/AN-04.pdf Picosecond Pulse Labs Application Note 4: TDR] | ||
Line 36: | Line 52: | ||
<gallery> | <gallery> | ||
1s2_front.jpg|front view | |||
1s2_left.jpg|left side view | |||
1s2_right.jpg|right side view | |||
1s2_bottom.jpg|bottom view | |||
Tek 1s2 e 1.jpg|time mode | Tek 1s2 e 1.jpg|time mode | ||
Tek 1s2 e 2.jpg|distance mode | Tek 1s2 e 2.jpg|distance mode | ||
Line 47: | Line 63: | ||
Tek 1s2 e 6.jpg | Tek 1s2 e 6.jpg | ||
Tek 1s2 e 7.jpg | Tek 1s2 e 7.jpg | ||
Tek 1s2 regulated ps.png|Internal Regulated Power Supply | |||
Tek 1s2 rear.jpg|rear | |||
</gallery> | </gallery> | ||
Line 54: | Line 72: | ||
[[Category:Time-domain reflectometers]] | [[Category:Time-domain reflectometers]] | ||
[[Category:500 series plugins]] | [[Category:500 series plugins]] | ||
[[Category:GR874]] |