Line 3: Line 3:
image=2440_1.jpg |
image=2440_1.jpg |
caption=Tektronix 2440  |
caption=Tektronix 2440  |
introduced=1988 |
introduced=1988 |  
discontinued=(?) |
discontinued=(?) |
summary=300 MHz, 500 MS/s 2-ch digital scope |
summary=300 MHz, 500 MS/s 2-ch digital scope |
manuals=
manuals=
* [http://w140.com/tek_2440_service.pdf Tektronix 2440 Manual (complete, B014064 and below) (PDF)]
* [http://w140.com/tek_2440_service.pdf Tektronix 2440 Manual (complete, B014064 and below) (PDF, OCR)]
* [http://w140.com/tek_2440_programmer_man.pdf Tektronix 2440 Programmer Manual (PDF, OCR)]
* [http://w140.com/tek_2440_programmer_man.pdf Tektronix 2440 Programmer Manual (PDF, OCR)]
* [http://w140.com/tek_2440.pdf Tektronix 2440 Manual (without schematics) (PDF)]
* [http://w140.com/tek_2440.pdf Tektronix 2440 Manual (without schematics) (PDF)]
* [[Media:Tek-2430A-2432A-2440-catalog.pdf|2430A/2432A/2440 catalog page (PDF)]]
* [[Media:070-6603-00 complete.pdf|Tektronix 2440 Manual (complete, B014064 and below) (PDF, OCR)]]
* [[Media:070-6603-00.pdf|Tektronix 2440 Manual (without schematics) (PDF, OCR, bad-OCR)]]
* [[Media:2430A-2432A-2440-catalog.pdf|2430A/2432A/2440 catalog page (PDF, OCR)]]
* [[Media:2430A-2432A-2440-catalog.pdf|2430A/2432A/2440 catalog page (PDF, OCR)]]
}}
}}
The '''Tektronix 2440''' is a two-channel 300 MHz digital sampling oscilloscope from the late 1980s.
The '''Tektronix 2440''' is a two-channel 300 MHz, 500 MS/s digital sampling oscilloscope from the late 1980s.
The maximum sample rate is 500 megasamples per second, per channel.
 
It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CCD output signals at 4 MS/s.


<blockquote>
<blockquote>
Line 26: Line 21:
</blockquote>
</blockquote>


{{MissingSpecs}}
{{BeginSpecs}}
 
{{Spec | Bandwidth | 300 MHz (Limit: 20, 100, Full) }}
'''Options'''
{{Spec | Sampling rate | 500 MSa/s each channel with simultaneous acquisition }}
{{Spec | Deflection | 2 mV/div to 5 V/div in 1-2-5 sequence }}
{{Spec | Auto Scale Factor | On-Screen Readout indicates probe tip deflection factors for ×1, ×10, ×100 and ×1000 probes }}
{{Spec | Input Impedance| 1 MΩ // 15 pF }}
{{Spec | Sweep Speeds | 2 ns/div to 5 s/div in 1-2-5 sequence (Modes A, A Intensified and B)}}
{{Spec | Auto Measure| Simultaneously measures up to four parameters on one or more displayed waveforms, updated at 3 Hz intervals or display a snapshot of 20 parameters of waveform target }}
{{Spec | Acquisition Modes| Normal, Envelope (glitch capture), Average }}
{{Spec | Display Modes| CH1, CH2, invert, add, multiply, X-Y }}
{{Spec | Cursors  | Volts, Time, Volts At Time, 1/Time, Slope }}
{{Spec | Functions | A Mode - Auto Level, Auto/Roll, Normal and Single Seq; B Mode - Trigger After Delay, Runs After Delay, 1K Record Length and more }}
{{Spec | Interface | GPIB }}
{{Spec | Probes    | Two passive [[P6137]] 400 MHz ×10 voltage probes, 1.3 m }}
{{Spec | Power    | 90 to 132 V<sub>AC</sub> or 80 to 250 V<sub>AC</sub>, 48 to 440 Hz; 160 W }}
{{Spec | Features  |
* Autoprobe Function With [[P6137]] Probe
* 300 MHz X-Y bandwidth at probe tip
* Automatic measurements
* extensive triggering capabilities
* Printer/Plotter Output
* On-screen HELP text describing every front-panel control
}}
{{Spec | Options |
* 03 - Word Recognizer ([[P6407]])
* 03 - Word Recognizer ([[P6407]])
* 05 - Video Waveform Measurement System
* 05 - Video Waveform Measurement System
* 11 - Two [[LEMO S-series connector|probe-power connectors]] on the rear panel
* 11 - Two [[LEMO S-series connector|probe-power connectors]] on the rear panel
}}
{{EndSpecs}}


==Links==
==Links==
* [https://hohlerde.org/rauch/elektronik/geraete/tek2440/index.en.html Tektronix 2440 internals & repair @ hohlerde.org]
* [https://hohlerde.org/rauch/en/elektronik/geraete/tek2440/  Tektronix 2440 internals & repair @ hohlerde.org]
 
==Internals==
The 2440 employs several processors − an MC68B09 main CPU controls the instrument, a custom 16-bit waveform processor performs signal computations, and an MC6805 microcontroller handles the front panel interface.  The GPIB Interface is implemented with a TMS9914 controller.
 
The power supply consists of a buck-type pre-regulator followed by a push-pull inverter and linear post-regulators (except for the +5 V digital supply rail).
 
CRT is a [[154-0850-01]].


==Pictures==
==Pictures==
Line 44: Line 69:
Tek 2440 5.jpg
Tek 2440 5.jpg
Tek 2440 rear2.jpg
Tek 2440 rear2.jpg
2440 1A.jpg
2440 1B.jpg
</gallery>
</gallery>



Revision as of 13:34, 12 March 2019

{{{manufacturer}}} 
300 MHz, 500 MS/s 2-ch digital scope
Tektronix 2440

Produced from 1988 to (?)

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix 2440 is a two-channel 300 MHz, 500 MS/s digital sampling oscilloscope from the late 1980s.

It uses CCDs (essentially, analog shift registers) to store input samples at high speed and a single 8 bit flash A/D converter to digitize the time-multiplexed CCD output signals at 4 MS/s.

WARNING:
Operation of the 2440 (and 2432 and 2430): A/D-converter and CCD hybrids may suffer irreparable damage within a few minutes if operated out of the metal enclosure without extra cooling.

Key Specifications

Bandwidth 300 MHz (Limit: 20, 100, Full)
Sampling rate 500 MSa/s each channel with simultaneous acquisition
Deflection 2 mV/div to 5 V/div in 1-2-5 sequence
Auto Scale Factor On-Screen Readout indicates probe tip deflection factors for ×1, ×10, ×100 and ×1000 probes
Input Impedance 1 MΩ // 15 pF
Sweep Speeds 2 ns/div to 5 s/div in 1-2-5 sequence (Modes A, A Intensified and B)
Auto Measure Simultaneously measures up to four parameters on one or more displayed waveforms, updated at 3 Hz intervals or display a snapshot of 20 parameters of waveform target
Acquisition Modes Normal, Envelope (glitch capture), Average
Display Modes CH1, CH2, invert, add, multiply, X-Y
Cursors Volts, Time, Volts At Time, 1/Time, Slope
Functions A Mode - Auto Level, Auto/Roll, Normal and Single Seq; B Mode - Trigger After Delay, Runs After Delay, 1K Record Length and more
Interface GPIB
Probes Two passive P6137 400 MHz ×10 voltage probes, 1.3 m
Power 90 to 132 VAC or 80 to 250 VAC, 48 to 440 Hz; 160 W
Features
  • Autoprobe Function With P6137 Probe
  • 300 MHz X-Y bandwidth at probe tip
  • Automatic measurements
  • extensive triggering capabilities
  • Printer/Plotter Output
  • On-screen HELP text describing every front-panel control
Options

Links

Internals

The 2440 employs several processors − an MC68B09 main CPU controls the instrument, a custom 16-bit waveform processor performs signal computations, and an MC6805 microcontroller handles the front panel interface. The GPIB Interface is implemented with a TMS9914 controller.

The power supply consists of a buck-type pre-regulator followed by a push-pull inverter and linear post-regulators (except for the +5 V digital supply rail).

CRT is a 154-0850-01.