Page values for "2755"

Jump to navigation Jump to search

"Specifications" values

21 rows are stored for this page
Parameter   Frequency
Value   
  • Coaxial Input Bands: 10 kHz to 21 GHz
  • External Waveguide mixers: 18 kHz to 325 GHz
Parameter   Frequency Span
Value   

100 Hz/div to 10 GHz/div in 1-2-5 sequence (plus 0 Hz and MAX)

Parameter   Resolution Bandwidth
Value   

6 dB bandwidths from 100 Hz to 3 MHz in decade steps

Parameter   RF Input impedance
Value   

50 Ω

Parameter   Maximum Safe Input Power
Value   
  • CW Power: 1 Watt (+30 dBm);
  • Pulse Power: 75 W peak (1 µS pulse width, 0.1% duty factor)
Parameter   External Mixer Input
Value   

50Ω for IF frequency inputs from external mixers covering 18 to 325 GHz

Parameter   RF Attenuator
Value   

0 dB to 60 dB, 10 dB steps

Parameter   Reference Level
Value   

-117 dBm to +40 dBm (+30 dBm maximum safe input)

Parameter   LO Output Power (2 to 6 GHz)
Value   

1st LO: +7.5 dBm to +15 dBm; 2nd LO: -22 dBm to +15 dBm; Must be terminated into 50 Ω when they are not connector to an external device

Parameter   Sweep Speed
Value   

5 s to 2 µs in 1, 2, 5 sequence (10 s/div in auto)

Parameter   Video Bandwidth
Value   

0.3 Hz to 30 kHz

Parameter   Triggering Modes
Value   

Internal, External, Line, Free Run

Parameter   Displayed Average Noise
Value   

−125 dBm

Parameter   Display Dynamic Range
Value   

90 dB

Parameter   Dynamic Range (Compression to noise)
Value   

134 dBm

Parameter   Calibrator (Cal out)
Value   

50Ω, -20 dBm ±0.3 dB at 100 MHz

Parameter   Included Accessories
Value   

Operator’s Manual; Programmer’s Manual; 6-ft, 50Ω coaxial cable, N-N; 18-inch, 50Ω coaxial cable, BNC-BNC; N male to BNC female adapter; rear connector shield; power cord, power cord clamp, and spare fuses; CRT filter set consisting of amber and gray light filters plus mesh filter

Parameter   Weight
Value   

27 kg (60 lbs)

Parameter   Power
Value   

90 − 132 VAC, 48 to 448 Hz; 180 – 250 VAC, 48 to 72 Hz. At 115 VAC, 60 Hz, 210 W max with all options

Parameter   Operating Atmospherics
Value   

Temperatures: 0 °C to +50 °C; Humidity: 95% below +30 °C, 75% above +30 °C, 45% above +40 °C; Altitude: Up to 3 km (10,000 ft)

Parameter   GPIB Characteristics
Value   

P Version: In accordance with IEEE Standard 488-1978 implemented as SH1, AH1, T5, L3, SR1, RL1, PP1, DC1, DT1, and C0; Non-P Version (Plotter Output): Implemented as SH1, AH0, T3, L0, SR0, RL0, PP0, DC0, DT0, and C0

"Products" values

1 row is stored for this page
Manufacturer   Tektronix
Model   2755AP
Class   Spectrum Analyzer
Description   

21 GHz Spectrum Analyzer

Introduced   1987
Discontinued   1993