{{BeginSpecs}}
{{Spec | Deflection | 10 μV/Div to 10 V/Div in 1–2–5 sequence }}
{{Spec | Deflection | 10 μV/Div to 10 V/Div in 1–2–5 sequence }}
{{Spec | Input impedance | 1 MΩ // 47 pF (either input) to 10MΩ switchable }}
{{Spec | Input impedance | 1 MΩ // 47 pF (Volts mode) or 10 MΩ (transducer mode) }}
{{Spec | Bandwidth | 1 MHz, LF limit switchable DC, 0.1 Hz to 10 kHz in six steps, HF limit switchable 100 Hz to 1 MHz in ×3/×10 steps, AC coupling 1.6 Hz }}
{{Spec | Bandwidth | 1 MHz, LF limit switchable DC, 0.1 Hz to 10 kHz in six steps, HF limit switchable 100 Hz to 1 MHz in ×3/×10 steps, AC coupling 1.6 Hz }}
{{Spec | Signal ranges | }}
{{Spec | Signal ranges | }}

Revision as of 06:00, 20 November 2018

Template:Plugin Sidebar 2 The Tektronix 3A10 is a differential amplifier plug-in for 560-series scopes.

It is intended for use with transducers and for that purpose it has a DC-coupled differential input and a variable voltage supply for powering the transducer. It also has a 6-pin input connector (Bendix MS3101A-10-6S) to directly connect to the 015-0169-00 Strain Gage Adapter.

The 3A10 circuit is partially based on the 1A7A's.

Key Specifications

Deflection 10 μV/Div to 10 V/Div in 1–2–5 sequence
Input impedance 1 MΩ // 47 pF (Volts mode) or 10 MΩ (transducer mode)
Bandwidth 1 MHz, LF limit switchable DC, 0.1 Hz to 10 kHz in six steps, HF limit switchable 100 Hz to 1 MHz in ×3/×10 steps, AC coupling 1.6 Hz
Signal ranges
Range Differential signal DC Offset Common mode
10 μV/Div to 10 mV/Div ±1 V ±1 V ±10 V
20 mV/Div to 0.1 V/Div ±10 V ±10 V ±100 V
0.2 V/Div to 1 V/Div ±100 V ±100 V ±500 V
2 V/Div to 10 V/Div ±1000 V ±1000 V ±500 V


Pictures