       9.  Next, check voltage at the base of Q4228 (on 465B-Early) schematic shows this should be about .7VDC.  With a "shorted" HV Multiplier, you will likely see 4.5-4.8VDC on the base of Q4228.
       9.  Next, check voltage at the base of Q4228 (on 465B-Early) schematic shows this should be about .7VDC.  With a "shorted" HV Multiplier, you will likely see 4.5-4.8VDC on the base of Q4228.
           a.  The regulation circuit is attempting to drive the circuit to the proper potential of about -2450VDC at TP4129 and 14-16kV at the CERT connector.  
           a.  The regulation circuit is attempting to drive T4015 to the proper potential of about -2450VDC at TP4129 and 14-16kV at the CERT connector.  
           b.  My 468 was showing only about -330VDC and a blank CRT.
           b.  My 468 was showing only about -330VDC and a blank CRT.
           c.  The HV Multiplier has likely shorted to ground and is demanding more current and voltage from T4015 than the transformer is able to deliver.   
           c.  The HV Multiplier has likely shorted to ground and is demanding more current and voltage from T4015 than the transformer is able to deliver.   