Page values for "7250"

Jump to navigation Jump to search

"Products" values

1 row is stored for this page
FieldField typeValue
ManufacturerPageTektronix
ModelString7250
ClassStringOscilloscope
DescriptionWikitext string

6 GHz Transient Digitizer

SeriesPage
IntroducedText1986
DesignersList of Page, delimiter: ;
DiscontinuedText1992(?)

"Specifications" values

13 rows are stored for this page
FieldField typeValue
ParameterStringBandwidth
ValueWikitext

6 GHz (measured as -3 dB @ 7 GHz, -6 dB @ 9.6 GHz, and -8.5 dB @ 14 GHz in Haas, Warman, van Ewijk, Rev. Sci. Instrum. 60 (3), March 1989)

FieldField typeValue
ParameterStringRise time
ValueWikitext

50 ps

FieldField typeValue
ParameterStringResolution
ValueWikitext

Vertical, 11 bits. The digitizer captures 512 samples in one shot.

FieldField typeValue
ParameterStringSweep
ValueWikitext

1 μs/Div to 50 ps/Div (1-2-5), equivalent to max. ~1 terasample per second

FieldField typeValue
ParameterStringSensitivity
ValueWikitext

fixed 500 mV/Div (no vertical amplifiers in the signal path)

FieldField typeValue
ParameterStringMax. input
ValueWikitext

2 kV spikes < 1 μs

FieldField typeValue
ParameterStringTriggering
ValueWikitext

External only, requires 50 ns pretrigger signal

FieldField typeValue
ParameterStringBandwidth
ValueWikitext

4.5 GHz

FieldField typeValue
ParameterStringRise time
ValueWikitext

≤75 ps

FieldField typeValue
ParameterStringDelay
ValueWikitext

55 ns

FieldField typeValue
ParameterStringAttenuation
ValueWikitext

4.5 dB

FieldField typeValue
ParameterStringInput / output
ValueWikitext

50 Ω N connectors

FieldField typeValue
ParameterStringTrigger pickoff
ValueWikitext

20 dB attenuation, 300 ps rise time, BNC connector