84: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 34: Line 34:
Tek 067-0523-00 4.jpg|067-0523-00
Tek 067-0523-00 4.jpg|067-0523-00
Tek 84 1.jpg|84
Tek 84 1.jpg|84
Tek 84 1.jpg|84
Tek 84 2.jpg|84
Tek 84 1.jpg|84
Tek 84 3.jpg|84
Tek 84 1.jpg|84
Tek 84 4.jpg|84


</gallery>
</gallery>

Revision as of 08:39, 17 February 2017

Template:Plugin Sidebar 2

The Tektronix Type 84 is a calibration fixture plug-in for 580-series scopes.

It was renamed the 067-0523-00 during production. The Type 84 provides a calibrated amplitude signal for adjusting gain, a fast-rise voltage step for adjusting the impulse response of the vertical amplifiers in the scope, and load resistors to test the power supply of the scope. The fast-rise pulses are produced by a mercury switch and are specified to have a repetition rate of 550 to 750 pulses per second, with a rise time that is "considerably less than the response time of the vertical amplifier in a 580-series oscilloscope."

Specifications

please add