==Internals==
==Internals==
Early production used a pair of [[5814]] tubes at the input while late used [[12AU7]]'s.
Instruments below serial number 3462 used a pair of [[5814]] tubes at the input; later production used [[12AU7]]'s.
These two tube types have different heater currents, and because the heaters are
These two tube types have different heater currents, and because the heaters are
series-wired, you cannot switch from one type to the other without circuit modification.
series-wired, you cannot switch from one type to the other without circuit modification.


==Specifications==
{{BeginSpecs}}
[[Category:Specifications needed]] ''please add''
{{Spec | Deflection | 1 mV/Div to 50 V/Div in 1–2–5-10-20-50 sequence and 1-10-100-1000 sequence }}
{{Spec | Input impedance | 1 MΩ // 47 pF (either input) }}
 
{{Spec | Bandwidth | 2 MHz with MV/CM MULTIPLIER set to 50;  300kHz with MV/CM MULTIPLIER set to 1 }}
{{Spec | CMRR | 10000:1 with MILLIVOLTS/CM set to 1, 1000:1 otherwise }}
{{Spec | Signal ranges | }}
{| class="wikitable" border=1 style="margin-left: 8em;" |
!Range
! Common mode dynamic range*
|-
| MILLIVOLTS/CM = 1 || ±2.5 V
|-
| MILLIVOLTS/CM = 10 || ±25 V
|-
| MILLIVOLTS/CM = 100 || ±250 V
|-
| MILLIVOLTS/CM = 1000 || ±500 V
|-
|}
* In practice, Type D behaves well up to about +/-15V if the CM signal is either a brief excursion or a constant.  If the CM varies slowly, DC shift in the input tubes will degrade CMRR.
{{EndSpecs}}