Page values for "DC504"
Jump to navigation
Jump to search
"Products" values
1 row is stored for this pageManufacturer | Tektronix |
Model | DC504 |
Class | Plug-in |
Description | 100 MHz, 5-digit frequency counter |
Series | TM500 system |
Introduced | 1975 |
Discontinued | 1993 |
"Specifications" values
14 rows are stored for this pageParameter | Frequency Range |
Value | 0 Hz (AC coupled: 10 Hz) to at least 80 MHz |
Parameter | Sensitivity (sine wave) |
Value | 20 mVRMS below 15 MHz, 35 mVRMS to 50 MHz, derated down to 175 mVRMS at 80 MHz |
Parameter | Input impedance |
Value | 1 MΩ // 20 pF |
Parameter | Trigger Level Range |
Value | −1.5 V to +1.5 V |
Parameter | Maximum Frequency Resolution |
Value | kHz ranges: 0.1 Hz; 1 Hz and 10 Hz (7 digit resolution possible in overflow kHz position); MHz ranges: 100 Hz, 1 kHz (9 digit resolution possible in overflow kHz position) |
Parameter | Maximum period resolution |
Value | ms ranges: 1 µs, 10 µs; s ranges: 10 ms, 1 ms, 0.1 ms |
Parameter | Trigger Error |
Value | 0.5% of one cycle at min. trigger sensitivity |
Parameter | Display Time |
Value | Variable ~0.1 s ... ~10 s and HOLD mode |
Parameter | Stability |
Value | 0°C to +50°C after 30 min warm up: 1×10-5 |
Parameter | Long Term Drift |
Value | 1×10-5 per month |
Parameter | Accuracy |
Value | Adjustable to within 1×10-7 |
Parameter | Stability |
Value | 0°C to +50°C after 30 min warm up: 1×10-7 |
Parameter | Long Term Drift |
Value | 1×10-7 per month |
Parameter | Accuracy |
Value | Adjustable to within 5×10-9 |