In many classic scopes, the high voltage transformer that supplies the cathode, blanking, and anode voltages has a tendency to break down.   
In many classic scopes, the '''high voltage transformer''' that supplies the cathode, blanking, and anode voltages has a tendency to break down.   
This is mostly because of aging of the potting material used for the high-voltage windings.   
This is mostly because of aging of the potting material used for the high-voltage windings.   
When this happens, the transformer will run very hot because dielectric loss is too high.  
When this happens, the transformer will run very hot because dielectric loss is too high.  
Line 10: Line 10:
This approach has been tested with success on a 556's HV transformer.
This approach has been tested with success on a 556's HV transformer.


The same HV transformer, part number 120-0308, is used in the [[543|543B]], [[544]], [[545|545B]], [[546]] and  [[547]] .
The same HV transformer, part number [[120-308]], is used in the [[543|543B]], [[544]], [[545|545B]], [[546]] and  [[547]] .
Chuck Harris, who frequents the Yahoo Tekscopes forum, has rebuilt several 120-0308 transformers with good results.
Chuck Harris, who frequents the Yahoo Tekscopes forum, has rebuilt several 120-308 transformers with good results.


<gallery>
<gallery>
KeepcoolHV.JPG | Alternative to replacement: cooling with a heat sink
KeepcoolHV.JPG | Alternative to replacement: cooling with a heat sink
120-466-0x.JPG | Two difference HV Transformer for [[561B]] & [[568]]. Brown potting (120-466-00) is bad, clear potting (120-0466-01) is ‎improved Variant to avoid hot running.
120-466-0x.JPG | Two different HV transformers for [[561B]] & [[568]]. Brown potting ([[120-0466-00]]) is bad, clear potting ([[120-0466-01]]) is ‎an improved variant to avoid the thermal problems.
</gallery>
</gallery>


[[Category:Repair issues]]
[[Category:Repair issues]]
[[Category:Inductors and transformers]]
[[Category:Inductors and transformers]]

Revision as of 02:19, 9 December 2018

In many classic scopes, the high voltage transformer that supplies the cathode, blanking, and anode voltages has a tendency to break down. This is mostly because of aging of the potting material used for the high-voltage windings. When this happens, the transformer will run very hot because dielectric loss is too high. It is a thermal runaway.

Currently, the only options when this happens are to find a similar transformer from another scope, or to get a rewound transformer.

A second option is to improve the cooling of the high voltage transformer. This approach has been tested with success on a 556's HV transformer.

The same HV transformer, part number 120-308, is used in the 543B, 544, 545B, 546 and 547 . Chuck Harris, who frequents the Yahoo Tekscopes forum, has rebuilt several 120-308 transformers with good results.