The P6015 is a high-voltage probe. It can be operated with or without it's freon filling.  Without the freon filling, the maximum voltage is reduced. The P6015 is a 1000X probe, and it has a compensating circuit
{{Probe Sidebar
at the scope end of the cable.
|manufacturer=Tektronix
|model=P6015
|summary=20(40) kV, ×1000 Passive Probe
|image=Tek_P6015_1.jpg
|caption=P6015
|for=1 MΩ scope inputs
|introduced=1963
|discontinued=1990
|manuals=
* [[Media:070-0373-01.pdf|P6015 manual 070-0373-01 (1963)]]
* [[Media:070-0373-02.pdf|P6015 manual 070-0373-02 (1985)]]
* [[Media:070-0373-02 rev 1990.pdf|P6015 manual 070-0373-02 (1990)]]


* [http://w140.com/p6015.pdf Tektronix P6015 Manual (early, PDF)]
* [[Media:Tek_p6007_p6009_p6015_cat.pdf|P6007 P6009 P6015 Catalog Description]]
* [http://w140.com/P6015_070-0373-02.pdf Tektronix P6015 Manual (late, PDF)]
}}
* [http://w140.com/P6015A_tek_original.pdf Tektronix P6015A Manual (PDF)]
The '''Tektronix P6015''' is a 1000× high-voltage probe usable up to 20 kV (40 kV peak pulse).


It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak), similar to the [[P6013]] predecessor model. 
The original filling material, [[wikipedia: 1,2-Dichlorotetrafluoroethane|Freon-114 aka Fluorocarbon 114 (R-114, CFC-114)]], is no longer available due to its [[wikipedia:Montreal Protocol|ozone-depleting properties]].  For this reason, it was replaced by the [[P6015A]] [[introduced in 1991|in 1991]], which uses a silicone based replacement.
The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants, which was available with [[PL-259]] or [[BNC]] plugs on the scope side (termination box part numbers 015-039/015-0039-00 and 015-049/015-0049-00, respectively).
{{BeginSpecs}}
{{Spec | Bandwidth          | 75 MHz (10' cable), 25 MHz (25' cable) }}
{{Spec | Risetime          | 4.67 ns (10' cable), 14 ns (25' cable) }}
{{Spec | Attenuation        | 1000× }}
{{Spec | Input impedance    | 100 MΩ // 3 pF }}
{{Spec | Compensation range | 15-50 pF }}
{{Spec | Max. input voltage |
* With dielectric filling: 20 kV DC or RMS to 100 kHz, pulse up to 40 kV (10%, < 0.1s)
* Without dielectric filling: 13 kV DC or RMS to 500 kHz, pulse up to 18 kV (10%, < 0.1s)
}}
{{EndSpecs}}
==Links==
* [[Patent US 3256484A|US Patent 3,256,484; ''High voltage test probe containing a part gas, part liquid dielectric fluid under pressure and having a transparent housing section for viewing the presence of the liquid therein'']].  Paul V Terry, 1962.
* [http://www.barrytech.com/tektronix/probes/tekp6015.html P6015 @ barrytech.com]
==Pictures==
<gallery>
<gallery>
File:P6015 BW.png
Tek P6015 box.jpg          | Tektronix P6015 in box
Tek P6015 box closed.jpg
Tek_P6015_1.jpg
Tek_P6015_2.jpg
Tek_P6015_3.jpg
Tek_P6015_4.jpg
P6015 BW.png
Tek p6015 schem.png        | schematic <br> old version
Tek P6015 derating.jpg      | Derating curves
P6015_Freon_Can.jpg        | Can of Freon Dielectric Fluid
Tek P6015 term box.jpg      | Termination box <br> new version
</gallery>
</gallery>
[[Category:1 MΩ Oscilloscope probes]]
[[Category:High voltage oscilloscope probes]]
387

edits