P6015

Revision as of 02:36, 22 April 2023 by Jaredcabot (talk | contribs) (Added datasheet)

The Tektronix P6015 is a 1000× high-voltage probe usable up to 20 kV (40 kV peak pulse).

It can be operated with or without dielectric filling but without this filling, the maximum voltage is reduced to 13 kV (18 kV peak), similar to the P6013 predecessor model.
The original filling material, Freon-114 aka Fluorocarbon 114 (R-114, CFC-114), is no longer available due to its ozone-depleting properties. The CFC-free P6015A successor, introduced in 1991, uses a silicone based replacement.

The P6015 has a compensating circuit at the scope end of the cable with multiple adjustable RC constants.

Key Specifications

Bandwidth 75 MHz (10' cable), 25 MHz (25' cable)
Risetime 4.67 ns (10' cable), 14 ns (25' cable)
Attenuation 1000×
Input impedance 100 MΩ // 3 pF
Compensation range 15-50 pF
Max. input voltage
  • With dielectric filling: 20 kV DC or RMS to 100 kHz, pulse up to 40 kV (10%, < 0.1s)
  • Without dielectric filling: 13 kV DC or RMS to 500 kHz, pulse up to 18 kV (10%, < 0.1s)

Links

Pictures